Manual

What to do if... ProLab 4000
200
ba75646e03 12/2012
Warning [TempErr]
Warning [ISEErr]
Burette does not dose
16.3 Conductivity measurement
Conductivity measuring
cell was not recognized
Error message, Error
(invalid calibration)
Cause Remedy
Temperature difference
between measurement and cali-
bration greater than 2 K.
Adjust the temperature of
the test sample
Cause Remedy
Electrode voltage outside cali-
brated range
Recalibrate
Cause Remedy
Address of the burette incor-
rectly set
Enter the address of the
burette on the meter
(see page 101)
Enter the address of the
burette on the burette
(see operating manual of the
burette)
RS232 interface of the burette
incorrectly set
Configure the RS232 inter-
face of the burette
(see page 29 and operating
manual of the burette)
Cause Remedy
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Unsuitable control standard Use control standard,
0.01 mol/l KCl