ProLab 4000 What to do if...
201
ba75646e03 12/2012
CalClock
with red frame
Cond temp?
16.4 General errors
Meter does not react to
keystroke
Time is lost
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Test sample does not have the
correct temperature (max.
± 2
°C temperature difference from
calibration temperature)
Adjust the temperature of the
test sample
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <On/Off> and
<SC> key simultaneously.
Cause Remedy
Buffer batteries empty Exchange buffer batteries
(see page 195)