Operating manual SensoLyt®MPP-A pH combination electrode ba75427defs01 04/2004 5
SensoLyt® MPP-A Application Application Measurements in normally loaded surface water and waste water. Installation Note Before installing the electrode in an armature, lubricate the O ring of the armature with the grease provided. Calibrating, measuring Note See operating manual of the pH meter. Note Take off the watering cap when you want to calibrate or measure. Identifying the electrode A series number is impressed on the pH electrode.
SensoLyt® MPP-A Cleaning Cleaning Contamination Cleaning Gross contamination at the diaphragm Brush off carefully using a soft toothbrush under running water Contamination at the membrane Rinse the electrode with water, then wipe it with a moist paper towel (do not rub) Fats, oils, protein-containing coatings and similar substances Remove with household washing-up liquid Accessories Buffer solutions Technical buffers Model Order no. pH 4.01 (1 liter) TEP 4 108 700 pH 7.
SensoLyt® MPP-A Technical data Technical data pH measuring range Temperature range Pressure range 4 ... 12 0 ... 50 °C 0 ... 106 Pa (10 bar) overpressure The SensoLyt® MPP-A electrode meets the requirements according to article 3(3) of the directive 97/23/EG ("pressure equipment directive").