User Manual

TetraCon 325 / TetraCon 325/C Cleaning
13
2 Cleaning
Caution
Before cleaning, disconnect the sensor from the instrument.
Cleaning A thorough cleaning is particularly recommended for measurements of low
conductivities.
After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.
Aging of the
conductivity
measuring cell
Normally, the conductivity measuring cell does not age. Special measuring
media (e.g. strong acids and lyes, organic solvents) or too high temperatures
shorten the operational lifetime considerably or damage the measuring cell.
The warranty does not cover cases where such conditions cause failure or
mechanical damage.
Disposal We recommend to dispose of the conductivity cell as electronic waste.
3 What to do if...
Contamination Cleaning procedure
Lime sediments Immerse in acetic acid for 5 minutes (volume
share = 10 %)
Fat/oil Clean with warm water that contains
washing-up liquid
Error symptom Cause Remedy
No temperature or con-
ductivity display
No connection between
measuring instrument and sensor
Cable defective
Check connection between
measuring instrument and sensor
Measurement delivers
implausible conductivity
values
Incorrect cell constant adjusted at
the measuring instrument
Measuring range exceeded
Contamination in the area of the
electrodes
Electrodes damaged
Check / correct the cell constant
Observe the application range
Clean the sensor (see section 2).
Return the sensor
Incorrect temperature
display
The temperature sensor was not
immersed deep enough in the
measuring solution
Temperature sensor defective
Observe the minimum immersion
depth
Return the sensor