Operating manual Turb 430 IR Turb 430 T Handheld Turbidimeter ba75507e02 12/2005
Turb 430 IR/T Accuracy when going to press The use of advanced technology and the high quality standard of our instruments are the result of continuous development. This may result in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
Turb 430 IR/T Contents 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.1 1.2 1.3 1.4 1.5 2 4.4 4.5 4.6 4.7 07/2006 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connecting the LabStation . . . . . . . . . . . . . . . . . . . . . . . Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 13 15 17 Operation . . . . . . . .
Contents Turb 430 IR/T 4.7.2 Configuring the RS232 interface . . . . . . . . . . . . 4.7.3 Selecting the output format of datasets . . . . . . . 4.7.4 Transmitting data . . . . . . . . . . . . . . . . . . . . . . . . 4.8 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.8.1 Resetting the system settings . . . . . . . . . . . . . . 4.8.2 Resetting turbidimeter settings . . . . . . . . . . . . . 4.9 Meter information . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Turb 430 IR/T Overview 1 Overview 1.1 General features The compact Turb 430 IR/T handheld precision turbidimeter enables you to carry out turbidity measurements quickly and reliably. The Turb 430 IR/T handheld meter provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
Overview Turb 430 IR/T 1.2 Keypad . Key functions Switch to the measured value display Start calibration . Open menus / confirm entries / start measurement Call up the Configuration menu (all settings are made here)
Turb 430 IR/T Overview Note Keys with an additional number printed on are assigned doubly. This enables to directly enter numbers in special menus. Thus, you can, for example, conveniently enter the date and time via the number keys. 1.3 Display The graphic display shows all information of the current measurement in the measured value display. The illumination enables to read the display even in the darkness. Example Turbidity 157.
Overview Turb 430 IR/T 1.5 LabStation (optional) With the LabStation, which is available as an accessory, you can conveniently use the Turb 430 IR/T in the laboratory. Laboratory operation with the LabStation enables the following additional functions: ! Line power operation is possible to save the batteries or accumulator pack ! The accumulator pack in the Turb 430 IR/T is automatically charged as soon as the meter is placed in the LabStation.
Turb 430 IR/T Safety 2 Safety This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual carefully before working with the meter. The operating manual must always be available within the vicinity of the meter. Target group The meter was developed for work in the field and in the laboratory.
Safety Turb 430 IR/T 2.1 Authorized use This meter is authorized exclusively for carrying out turbidity measurements in the field and laboratory. The technical specifications as given in chapter 7 TECHNICAL DATA must be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized. 2.
Turb 430 IR/T Safety Safe operation It is the responsibility of the operator to continuously observe the overall technical condition (externally recognizable deficits and damage as well as alterations to the operational behavior) of the meter.
Safety 12 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T Commissioning 3 Commissioning 3.1 Scope of delivery ! Handheld turbidimeter Turb 430 IR or Turb 430 T ! 4 batteries, 1.
Commissioning Turb 430 IR/T Note The accumulator pack should not be completely discharged. If you do not operate the instrument for a longer period of time you should charge the accumulator pack every six months. Automatic switchoff function The meter has an automatic switch-off function in order to save the batteries or accumulator pack (see section 4.4).
Turb 430 IR/T Commissioning 3.3 Connecting the LabStation Note The LabStation is available as an accessory (see section 8.1). In order to use the functions of the LabStation for operation in the laboratory, connect the LabStation and place the Turb 430 IR/T in the LabStation. Connecting the LabStation (optional) 5 4 3 2 1 ba75507e03 07/2006 1 If necessary, replace the Euro plug (1) on the power pack (2) by the country-specific plug suitable for your country.
Commissioning Turb 430 IR/T LabStation 16 ba75507e03 07/2006
Turb 430 IR/T Commissioning 3.4 Initial commissioning Perform the following activities: ! For – accumulator operation: insert the accumulator pack (see section 5.1.2) – line power operation and charging the accumulator pack: connect the power pack (see section 3.2) – operation with LabStation: connect the LabStation and place the Turb 430 IR/T in the LabStation (see section 3.3) ! Switch on the meter (see section 4.1) ! Set the language as necessary (see section 4.3.
Commissioning 18 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T Operation Switching on 4 Operation 4.1 Switching on the meter Press the key. The Start menu appears for 30 seconds. The status line indicates the meter designation and the version number of the software. Start Turbidity i Turb 430 IR V 0.24 Note in das Menü Start gelangen Sie bei eingeschaltetem Gerät durch ggf. mehrfaches Drücken der Taste . After a few seconds the meter automatically switches to the measuring mode.
Operation Display illumination with battery-powered operation Turb 430 IR/T During operation with batteries or accumulator pack the meter automatically switches off the display illumination if no key is pressed for 30 seconds. The illumination is switched on again with the next keystroke. 4.2 Inserting a cell To be able to insert cells in the Turb 430 IR/T, the cell shaft has to be prepared to take in a cell. 1 Push the dust cover (1) upward. The cell shaft for 28 mm cells is open.
Turb 430 IR/T Operation 4.3 General operating principles This section contains basic information on the operation of the Turb 430 IR/T. Operating elements, display Operating modes, navigation An overview of the operating elements and the display is given in section 1.2 and section 1.3. An overview of the operating modes of the Turb 430 IR/T and the navigation through menus and functions can be found in section 4.3.1 and section 4.3.2. 4.3.
Operation Turb 430 IR/T Configuration Turbidity System Info ! Settings Settings are indicated by a colon. The current setting is displayed on the right-hand side. With , the selection of the possible settings is opened. Subsequently, the setting can be changed with <▲> <▼> and. Example: System Language: Beep: Deutsch Off Illumination: Contrast: On 48 % Temperature unit: °C Switchoff time: 30 min ! Functions Functions are designated by the name of the function.
Turb 430 IR/T Operation Turbidity \ Calibration i Insert standard 1000 FNU/NTU i Press and hold i Align sample The i indicates info texts, e.g. messages, notes or instructions Note The principles of navigation are explained in the two following sections by reference of examples: ! Setting the language (section 4.3.3) ! Setting the date and time (section 4.3.4).
Operation Turb 430 IR/T 4.3.3 Navigation example 1: setting the language Note The following example describes in the language of the country how to set the language. On delivery, English is set as the language in the Turb 430 IR/T. During initial commissioning, the language is set in the menu, Configuration / System / Language. 1 In the measured value display: Open the Configuration menu with
Turb 430 IR/T Operation 8 4.3.4 To make further settings, switch to the next higher menu level with . Navigation example 2: setting the date and time The meter has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well. Numerals are generally entered via the number keys.
Operation Turb 430 IR/T 3 Select and confirm the Time menu with <▲> <▼> and . A display for the entry of numerals with the number keys opens up. Time 14:53:40 4 Enter the time using the number keys. The digit to be changed is displayed underlined. Note In the case of wrong entries, you can cancel the procedure with . After canceling with , it is possible to enter all digits once again. The new digits are only taken over by confirming with .
Turb 430 IR/T Operation System Language Deutsch English Français Español Measured value memory Display RS232 download Data filter Filter ID Date Delete i 4 of 1000 occupied i Filter: No filter Display Illumination Auto off On Off Contrast 0 ... 100 % Brightness 0 ... 100 % Baud rate 1200, 2400, 4800, 9600, 19200 Output format ASCII CSV Time hh:mm:ss Reset Interface Continue ... / Date/time Date Date format Continue ...
Operation Turb 430 IR/T 4.
Turb 430 IR/T Operation Menu item Setting Description Continue ... / Switchoff time 10, 20, 30, 40, 50 min, 1, 2, 3, 4, 5, 10, 15, 20, 24 h The automatic switchoff function switches the meter off if no entry is made for a specified period of time (Switchoff time). This saves the batteries or accumulator pack. Continue ... / Beep On Off Switch on/off the beep on keystroke 4.4.
Operation Settings/functions Turb 430 IR/T Menu item Setting/ function Description Display - Displays in pages all measurement datasets that correspond to the filter settings. Further options: ! Scroll through the datasets with <▲> <▼>. ! Output the displayed dataset to the interface with . ! Quit the display with . RS232 download - Downloads to the interface all measurement datasets that correspond to the filter settings. The download is ordered according to the date and time.
Turb 430 IR/T Operation 4.4.2 Display In the Configuration / System / Display menu, you set the display features: ! Switching on/off the display illumination (Illumination) ! Display contrast (Contrast) The settings are given in the menu, Configuration / System / Display. Move to the Configuration menu with the
Operation Turb 430 IR/T 4.4.4 Date/time In the Configuration / System / Continue ... / Date/time menu, you set the system clock: ! Current time (Time) ! Current date (Date) ! Format of the date display (Date format) The settings are given in the menu, Configuration / System / Continue ... / Date/time. Move to the Configuration menu with the
Turb 430 IR/T Operation Venting the sample Avoiding or removing air bubbles 4.5 Turbidity 4.5.1 General information Air bubbles in the sample affect the measuring result to a massive extent because they have a large scattering effect on the incident light. Larger air bubbles cause sudden changes in the measured values whereas smaller air bubbles are recorded by the instrument as turbidity.
Operation Turb 430 IR/T 4 Marking a cell Release the key. Measurement starts. The measured value is displayed. To be able to quickly bring a cell into the optimum position, it is helpful to mark the optimum position of the cell once it is determined. This shortens each measurement or calibration procedure with this cell considerably. The marking can, e. g., be done on a label on the cap of the cell. 5 4.5.3 Mark the optimum position of the cell.
Turb 430 IR/T Operation 6 Align the cell: ! Marked cell – Align the marking on the cell cap with the marking on the cell shaft. – Press and for a short time hold the key until the measured value is displayed. ! Unmarked cell (see page 33) – Press and hold the key. – Slowly and in small steps rotate the cell by one complete rotation (by 360 °). After each step wait for a short time until the displayed measured value is stable.
Operation Turb 430 IR/T 4.5.4 Overview Settings for turbidity measurements For turbidity measurements, the following settings are possible in the menu, Configuration / Turbidity: ! Calibration record (display, print) ! Entering the Calibration interval ! Reset Settings/functions The settings are given in the menu, Configuration / Turbidity. Move to the Configuration menu with the
Turb 430 IR/T Operation Calibration record Displaying and downloading calibration data to interface At the end of each calibration procedure a calibration info (i symbol) and the calibration record is displayed. You can view the data of the last calibration on the display. Subsequently, you can download the displayed calibration data to the interface, e. g. to a printer or PC, with the key.
Operation Turb 430 IR/T 3 Align the cell: ! Marked cell: – Align the marking on the cell cap with the marking on the cell shaft. – Press and hold the key until the measured value is displayed. ! Unmarked cell (see page 33) – Press and hold the key. – Slowly and in small steps turn the cell by one complete rotation (by 360 °). – After each step wait for a short time until the displayed measured value is stable.
Turb 430 IR/T Operation Turbidity i Insert sample i Press and hold i Align sample Note If i Calibration error! was displayed as the calibration result, a note appears on the display to recalibrate before measuring. Should a valid calibration not be possible the meter also offers to continue measuring with the last valid calibration data.
Operation Turb 430 IR/T 4.6 Saving The meter has 1000 memory locations for measurement datsets. You can transmit measured values (datasets) to the data memory with the key. Each storage process transmits the current dataset to the interface at the same time. The number of memory locations that are still free is displayed in the Store menu. The number of memory locations that are occupied is displayed in the System Measured value memory menu.
Turb 430 IR/T Operation If the memory is full You can erase the entire memory (see section 4.6.5), or overwrite the oldest dataset with the next storing procedure. A security prompt appears before a dataset is overwritten. 4.6.2 Filtering measurement datsets The functions to display and download stored measurement datsets (see section 4.4.1) refer to all stored measurement datsets that correspond to the specified filter criteria.
Operation Turb 430 IR/T 4.6.3 Displaying measurement datsets You can read out stored datasets to the display. Only those datasets are displayed that correspond to the selected filter criteria (see section 4.6.2). Start reading out the data to the display in the menu, Configuration / System / Measured value memory / Display. Representation of a dataset 02.02.2005 11:24:16 ID: 1 16.80 FNU/NTU i Scroll with ▲ ▼ Further datasets that correspond to the filter criteria are displayed with the <▲> <▼> keys.
Turb 430 IR/T Operation 4.6.5 Erasing stored measurement datasets You can erase the stored measurement datsets altogether if you do no longer need them. Erasing all measurement datsets is done in the menu, Configuration / System / Measured value memory / Delete. Note Erasing individual datasets is not possible. If all memory locations are occupied, however, it is possible to overwrite the oldest dataset at a time. A security prompt appears before a dataset is overwritten. 4.
Operation Turb 430 IR/T 4.7.1 Connecting a PC/external printer Connect the interface to the devices via the AK540/B (PC) or AK540/S (external printer) cable. Caution The RS232 interface is not galvanically isolated. When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
Turb 430 IR/T Operation 4.7.3 Selecting the output format of datasets For downloading data to the interface you can select the output format. It is selected in the menu, Configuration / System / Interface / Output format. The ASCII output format delivers formatted datasets. The CSV output format delivers datasets separated by ";". Output format, ASCII Turb 430 IR Ser. no. 12345678 31.10.04 09:56:20 ID: 1 16.01 FNU/NTU ________________________________________ Turb 430 IR Ser. no. 12345678 31.10.
Operation Turb 430 IR/T 1 Output format, CSV 2 3 4 5 6 15.01.05;15:06:49;1;26.5;FNU/NTU;VALID; 15.01.05;16:06:49;1;16.
Turb 430 IR/T Operation 4.7.4 Transmitting data The following table shows which data are transmitted to the interface in which way: Data Operation / description Current measured value ! Press . Stored measured values ! Simultaneously with every manual storage process. ! Display stored dataset and press . ! All datasets according to the filter criteria via the RS232 download function (see section 4.6.2).
Operation Turb 430 IR/T 4.8 Reset You can reset (initialize) all system and measurement settings. 4.8.1 Resetting the system settings With the System / Reset function, all resettable settings are reset. ! Settings for Turbidity (see section 4.8.2) ! System settings System setting Default settings Baud rate 4800 Baud Output format ASCII Illumination Auto off Contrast 50 % Brightness 50 % Switchoff time 30 min Beep On 4.8.
Turb 430 IR/T Operation 4.9 Meter information The following meter information is listed in the Configuration / Info menu: ! Model designation ! Software version ! Series number of the meter Info i Model: Turb 430 IR i Software: V 0.35 i Ser. no.: 04280003 4.10 Software update With a software update you obtain the current instrument software (see appendix). The current software version can be found on the Internet under www.WTW.com.
Operation 50 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T Maintenance, cleaning, disposal 5 Maintenance, cleaning, disposal 5.1 Maintenance The meter is almost maintenance-free. The only maintenance task is replacing the batteries or accumulator pack. 5.1.1 Inserting/exchanging the batteries Caution Make sure that the poles of the batteries are the right way round. The ± signs on the batteries must correspond to the ± signs in the battery compartment.
Maintenance, cleaning, disposal Turb 430 IR/T 5.1.2 Retrofitting the accumulator pack Caution Use original WTW accumulator packs only. Together with the power pack the accumulator pack is available as an accessory (see section 8.1). 1 2 3 1 Open the battery compartment: – Unscrew the two screws (1) on the underside of the meter, – Remove the lid of the battery compartment (2). 52 2 If necessary, take four old batteries out of the battery compartment.
Turb 430 IR/T Maintenance, cleaning, disposal 5.2 Cleaning Occasionally wipe the outside of the meter with a damp, lint-free cloth. Disinfect the housing with isopropanol as required. Caution The housing components are made out of synthetic materials (polyurethane, ABS and PMMA). Thus, avoid contact with acetone and similar detergents that contain solvents. Remove any splashes immediately. 5.2.1 Cleaning the cell shaft If liquid is in the cell shaft (e.g.
Maintenance, cleaning, disposal Turb 430 IR/T 5.3 Disposal Packing This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the instrument against damage during transport. Batteries/accumulator pack Remove the batteries or accumulator pack from the meter (see section 5.1). Dispose of the batteries or accumulator pack at a suitable facility according to local legal requirements.
Turb 430 IR/T What to do if... Display, LoBat Instrument does not react to keystroke 6 What to do if... 6.1 General errors Cause Remedy – The batteries or accumulator pack are largely depleted – Insert new batteries Cause Remedy – Software error – Processor reset: Press the and key simultaneously. – Operating condition undefined or EMC load unallowed Error message, Error 0, 8, 16, 16384 – Charge the accumulator pack (see section 3.
What to do if...
Turb 430 IR/T Technical data Dimensions Weight Mechanical structure 7 Technical data 7.1 General data 7.1.1 Turb 430 IR/T approx. 236 x 86 x 117 mm approx. 0.6 kg (without batteries) Type of protection IP 67 Electrical safety Protective class III Test certificates cETLus, CE, FCC Ambient conditions Allowable relative humidity Power supply Serial interface ba75507e03 07/2006 Storage - 25 °C ... + 65 °C Operation 0 °C ...
Technical data Turb 430 IR/T Guidelines and norms used EMC EC guideline 89/336/EEC EN 61326-1/A3:2003 FCC Class A Instrument safety EC guideline 73/23/EEC EN 61010-1 :2001 UL STD 61010-1 CAN/CSA-C22.2 No. 61010-1 Climatic class VDI/VDE 3540 IP protection EN 60529:1991 FCC Class A Equipment Statement Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
Turb 430 IR/T Technical data Measuring principle Light source Measuring range Resolution Accuracy 7.2 Turbidity 7.2.1 Turb 430 IR Nephelometric measurement according to DIN EN ISO 7027 Infrared LED 0.01 ... 1100 FNU/NTU in the range 0,01 ... 9,99 max. 0.01 FNU/NTU in the range 10,0 ... 99,9 max. 0.1 FNU/NTU in the range 100 ... 1100 max. 1 FNU/NTU in the range 0 ... 1100 FNU/NTU ± 2 % of the measured value or ± 0.
Technical data 60 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T Accessories, options 8 Accessories, options 8.1 WTW accessories Description Model Order no. LabStation pHotoFlex LS 251 301 Accumulator with Turb 430 IR/T power pack pHotoFlex BB 251 300 3 empty cuvettes, 28 x 60 mm LKS28-Set 251 302 Calibration standard kit for Turb 430 IR Kal.Kit Turb 430 IR 600 560 Calibration standard kit for Turb 430 T Kal.
Accessories, options Thermoprinter Turb 430 IR/T You can connect the P3001 to the Turb 430 IR/T in the following ways: Description Model Order no.
Turb 430 IR/T Lists 9 Lists This chapter provides additional information and orientation aids. Abbreviations The list of abbreviations explains the indicators and the abbreviations that appear on the display and in the manual. Specialist terms The glossary briefly explains the meaning of the specialist terms. However, terms that should already be familiar to the target group are not described here.
Lists Turb 430 IR/T Abbreviations 64 Cal Calibration d Day h Hour j Year LoBat Batteries almost empty (Low battery) m Month s Second S Slope (internat. k) SELV Safety Extra Low Voltage Slp.
Turb 430 IR/T Lists Glossary Adjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance. Calibration Comparing the value from a measuring system (e. g. the displayed value) to the correct value or a value that is regarded as correct.
Lists 66 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T 10 Index Index A M Accumulator pack Charging time .....................................13 Air ..........................................................33 Aligning and marking a cell .......................33 Authorized use .........................................10 Automatic switchoff function ................14, 19 Measured value display ........................... 21 Measured value memory ......................... 40 Measurement dataset ..............................
Index 68 Turb 430 IR/T ba75507e03 07/2006
Turb 430 IR/T Appendix: Firmware update General information With the "Firmware Update Turb430" program you can update the firmware of the Turb 430 IR/T to the latest version with the aid of a Personal Computer. A free serial interface (COM port) on your PC and an interface cable is required for this (see chapter 8 ACCESSORIES, OPTIONS). Program installation Program start Install the firmware update program on your PC with the "Turb430_Vx_yy_English.exe" installation program.
Turb 430 IR/T Appendix: Turbidity values under 1 FNU/NTU With turbidity values (under 1 FNU/NTU), the measured value is strongly influenced by the cell and its alignment. In order to increase measuring accuracy with turbidity values under 1 FNU/NTU, calibration in the 0.02 FNU/NTU standard and later measurement should be carried out in the same cell. For calibration in the standards 10.0 and 1000 FNU/NTU follow the instructions on the display.
Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: Fax: E-Mail: Internet: +49 (0) 881 183-0 +49 (0) 881 183-100 +49 (0) 881 183-420 Info@WTW.com http://www.WTW.