User guide

Operation Turb 430 IR/T
38
ba75507e03 07/2006
Note
Before measuring the third calibration standard of 0.02 FNU/NTU you
can exit the calibration with <ESC> at any time.
The new calibration data is discarded. The old calibration data is further
used.
3 Align the cell:
! Marked cell:
Align the marking on the cell cap with the marking on the
cell shaft.
Press and hold the <START/ENTER> key until the
measured value is displayed.
! Unmarked cell (see page 33)
Press and hold the <START/ENTER> key.
Slowly and in small steps turn the cell by one complete
rotation (by 360 °).
After each step wait for a short time until the displayed
measured value is stable.
Turn the cell back to the position with the lowest
measured value.
4 Release the <START/ENTER> key.
Measurement of the calibration standard begins.
5 Repeat the steps 4 - 6 with the calibration standards, 10.0 FNU/
NTU and 0.02 FNU/NTU.
After measuring the 0.02 FNU/NTU calibration standard, the
calibration result is displayed.
Calibration is completed.
6 Confirm the calibration result with <START/ENTER>.
The calibration record is displayed.
7 Confirm the calibration record with <START/ENTER>.
The display shows instructions for the first measurement.
i Turb. = 1000 FNU/NTU
i Start calibration
by releasing <START>
Turbidity \ Calibration