TX80-Series •TX81 •TX82 ISO 9 0 0 1 : 2 0 0 8 CERTIFIED COMPANY TX80-SERIES FLOW SENSOR INSTRUCTIONS FLOW SENSOR INSTRUCTIONS
TABLE OF CONTENTS General Information Specifications, Features........................................................................................................................................Page 1 Installation Distorted Flows, Fitting Installation, Meter Installation, Positioning the Meter.................................................Page 2 Straight Pipe Recommendations..........................................................................................................................
GENERAL INFORMATION, SPECIFICATIONS, FEATURES GENERAL INFORMATION The TX80-Series are insertion turbine meters designed for use in 1-1/2" to 8" pipe. High-quality jewel bearings and precision shafts ensure long life and low friction. Available in 316 stainless steel, brass and PVC, sensor bodies are machined from solid rod for maximum low-flow performance. The TX80-Series use special fittings that ensure ease of installation and correct depth setting in the pipe.
INSTALLATION DISTORTED FLOWS Distorted flow profile tings (size 3” and above) require a hole to be cut in the pipe. The recommended hole size is 1-3/4”. Faster flow causes meter to read high FLOW 10X Diameter Minimum (See Below) Fitting Installation. TX80-Series meters require special fittings. The meter fitting must first be installed in the pipeline.
INSTALLATION STRAIGHT PIPE RECOMMENDATIONS (X = diameter) 10X 5X Reduced Pipe 10X Two Elbows In Plane Two Elbows, Out Of Plane 5X 20X 5X 20X 5X Expanded Pipe 30X Spiral Flow Propeller Meter 50X Swirling Flow Partially Open Butterfly Valve Page 3
INSTALLATION FULL PIPE RECOMMENDATIONS Better Installation: Ensures full pipe Possible Problem: Allows air pockets to form at sensor Possible Problem: Post-valve cavitation can create air pocket Better Installation: Keeps pipe full at sensor Possible Problem: Air can be trapped Better Installation: Allows air to bleed off Caution: These flow sensors are not recommended for installation downstream of the boiler feedwater pump where installation fault may expose the flow sensor to boiler pressure and t
CONNECTIONS DIAGRAMS FT415 Standard Connections Electronic Metering Pump Battery Type: Lithium "C, 3V, replaceable PULSE SCALED Caution: Do not apply external power to the FT415.
OPERATION Modularity. Any of the TX80-Series flow sensors can have electronic modules mounted directly to them, either at the time of order or by adding an adaptor kit in the field (part number MK10). Alternatively, the flow sensor and indicator or transmitter can be installed remote from each other. The FT420 is an indicating transmitter (rate, total, 4-20 mA output), and the AO55 is a blind 4-20 mA transmitter. For data logging, add the DL76.
MAINTENANCE Caution: Never remove the u-clip retainer when the pipe is under pressure. Always remove pressure from the pipe before attempting to remove the meter. Removal under pressure may result in damage or serious injury. Rotor Replacement. It is unusual for a rotor to require replacement due to damage sustained in normal service. More commonly, the meter is dropped while it is out of the pipe. Another reason for rotor replacement is shaft wear after long service. Rotors are easily field-replaced.
PARTS EXPLOSION 4 5 6 1 7 2 3 8 7 9 10 11 TX80-SERIES PARTS LISTING 12 13 1 Upper Housing 26181 2 Gasket 26211 3 Lower Housing 29930 4 Housing Screw 26229 5 Plug, Steel 26073 6 Plug, Plastic 26079 7 Strain Relief 07655 8 Sensor Retaining Screw 25321 Pick Up, Standard Low Power or Pick Up, Micro Power (FT415/DL76) O-Ring, EPDM 10 or O-Ring, Viton Rotor Repair Kit 11/12 (Polypro/Tungsten Carbide) 9 12 13 Page 8 26310 29953 25081 16455 33087 Rotor Assembly 33015 (Polypro
TROUBLESHOOTING Problem Probable Cause Try... No signal after installation Insufficient flow Reduce pipe size if possible, use different type of sensor Bad connections to control electronics Reconnect if necessary Incompatible control Use different power supply Add pull up resistor, (See pg. 5) Damaged or missing rotor Obtain new rotor and replace (See pg.
S e am e tri c s In c o r po r at e d • 1 9 0 2 6 72n d A ven ue South • Ken t, Washi n g ton 98032 • USA LT-65200048-100211 ( P ) 2 5 3 . 8 7 2 . 0 2 8 4 • ( F ) 2 5 3 . 8 7 2. 0285 • 1. 800. 975. 8153 • w w w . seametr i cs.