WMP-Series WMP-SERIES PLASTIC-BODIED MAGMETER INSTRUCTIONS PLASTIC-BODIED MAGMETER INSTRUCTIONS 3 inch 2 inch ISO 9 0 0 1 : 2 0 0 8 CERTIFIED COMPANY 1 inch
GENERAL INFORMATION and FEATURES GENERAL INFORMATION The WMP-Series is a full-bore, plastic-bodied electromagnetic flow meter designed for flow and usage monitoring applications in 1, 2 and 3 inch pipe. The polypropylene flow tube offers corrosion resistance to a wide range of chemicals. Its light weight and easy to install or remove from the pipe for inspection. With no moving parts, the magmeter permits unobstructed flow, minimizing flow disturbances and hence, straight pipe requirements.
SPECIFICATIONS* Pipe Size 1, 2 or 3 inch full port Fittings 1 inch NPTF, 2 or 3 inch flange clamps with 2 or 3 inch NPTF fitting kit Pressure 150 psi or 10.3 bar working pressure @ 70˚ F Operating Temperature Range 10˚ to 130˚ F (-12˚ to 54˚ C) operating, -40˚ to 176˚ F (-40˚ to 80˚ C) non-operating Accuracy +/-1% of reading (between 10% and 100% of max flow) +/-3% of reading (between cutoff and 10% of max flow) Flow Range Minimum Maximum 1 inch: 2.3 GPM (.145 LPS) 2 inch: 6 GPM (.
FLOW RANGE, INSTALLATION FLOW RANGE Minimum Maximum Gal/Min 2.3 110 1” Liter/Sec .145 6.94 Gal/Min 6 300 INSTALLATION Piping Conditions. Installing the meter with a length of straight pipe at least two times the diameter upstream and one diameters downstream is highly recommended. Some piping conditions require more than this. See chart for recommendations. Step 1. Position gasket at either end 2” Liter/Sec .38 18.9 Gal/Min 14 670 3” Liter/Sec .88 42.3 End Connections.
STRAIGHT and FULL PIPE RECOMMENDATIONS (X = diameter) Reduced Pipe Expanded Pipe 1X 2X 1X 5X WMP WMP Swirling Flow: Partially Open Butterfly Valve Elbows 1X 1X 2X 5X WMP WMP Recommended: Keep pipe full at meter for accuracy WMP WMP Not Ideal: Allows air pockets to form at meter WMP WMP Recommended: Keeps pipe full at meter for accuracy Not Ideal: Post-valve cavitation can create air pocket WMP WMP Recommended: Allows air to bleed off Not Ideal: Air can be trapped 4
ELECTRICAL CONNECTIONS, CABLE CONNECTIONS and OPERATION ELECTRICAL CONNECTIONS WMP104. The WMP104 is battery-powered totally self-contained and does not require any electrical connections (there is no output on the WMP104 model). CABLE CONNECTIONS Gray: Pin 5 = Ground White: Pin 4 = (–) Pulse Blue: Pin 3 = (+) Pulse Black: Pin 2 = (–) Power Brown: Pin 1 = (+) Power Shield Drain Wire OPERATION Grounding (WMP101).
MAINTENANCE, REPAIR and TROUBLE SHOOTING MAINTENANCE AND REPAIR There are no user-serviceable parts in the WMP-Series meters except the batteries in the WMP104. Battery Replacement. When the “low bat” indicator appears, the batteries should be changed. Either six each AA alkaline cells are required, or two each C lithium cells with integrated wiring harness, depending on power option ordered with meter. To change the batteries, first remove the four screws which hold the top cover in place.
S e am e t r i c s In c o r po r at e d • 1 9 0 2 6 72n d A ven ue South • Ken t, W ashi n g ton 98032 • USA ( P ) 2 5 3 . 8 7 2 . 0 2 8 4 • ( F ) 2 5 3 . 8 7 2. 0285 • 1. 800. 975. 8153 • w w w . seametr i cs.