WATER BATH BM110 BM210 First Edition ● Thank you for your Yamato Scientific BM series Water Bath purchase. ● For proper use of this unit, please read the instruction manual and warranty thoroughly before operation. Keep both for any future references. Read and apprehend important warnings in this instruction manual prior to use.
Table of Contents 1. Specifications..................................................................................................................................... 1 2. Safety Information .......................................................................................................................... 2-5 Safety Symbols…………………………………………………………………………..2 Safety Precautions…………………………………………………………………….3-4 Hazardous Materials…………………………………………………………………….5 3. Identification of Parts .....................
1. Specifications Model BM110 BM210 Temp. range during operation Room temp. +5℃ ~ 95℃ *1 Temp. setting range 0℃ ~ 100℃ Accuracy of temp.
2. Safety Information Safety Symbols Safety Information This instruction manual and our products apply various indications for safety. Ignoring these indications can cause such situations as listed below. Read and understand the following warning and caution signs in this manual prior to use. WARNING Indicates the possibility of serious or fatal injury (Note 1). CAUTION Indicates the possibility of injury (Note 2) or damage (Note 3) to the equipment.
Safety Information Safety Precautions Do not use this unit for any purpose other than its intended use, described in this manual. Do not use this unit in flammable or explosive gas environments. This unit is not explosion proof. Never use this unit in flammable or explosive gas environments. Never fail to ground the unit. This unit uses a 3-conductor power cord (including ground wire). Be sure to ground the unit for safety. Do not use this unit if a malfunction occurs.
Safety Information Safety Precautions If it begins to thunder… In the event of electrical storm turn off the main power. Neglecting this procedure can result in fire, electric shock or other troubles due to thunderbolts. In case of power failure… When the power is restored after a power failure, the unit resumes operation at the temperature previously set as if you turned off the unit. Indication of temperature and operational range. This unit is not equipped with a stirring function.
Safety Information Hazardous Materials Hazardous material are listed below. Never use these materials as samples or heat media. Nitroglycol, Nitroglycerin, Nitrocellulose, and other explosive nitric esters. Explosive Explosive Substance Combustible Substance Oxidant Trinitrobenzene, Trinitrotoluene, Picric acid, and other explosive nitro compounds. Peracetic acid, Methyl ethyl ketone peroxide, Benzoyl peroxide, and other organic peroxides.
3.
Identification of Parts BM210 BM210 Thermometer Power Switch Drain Plug Power Lamp Temperature Dial Fuse Power Cable Heater Thermometer Holder Sensor 7
Identification of Parts Sample : Combined with RE210 Combination with RE210 Drawing 1 Drawing 2 According to the position or space to set the bath and the body, the arm jack can be set in the front (drawing 1)or on the right side (drawing 2) as shown above.
Identification of Parts Sample : Combined with RE510 Combination with RE510 9
4. How to Operate How to operate Power Lamp Temperature Dial 1. Pour adequate water (3000ml for BM110, 5000ml for BM210) into the tank, and place a thermometer in the holder. Use distilled water. Deionized water should not be used. 2. Connect the unit securely to a power source with AC230V. Always use a grounded outlet. Set the temperature dial to the desired temperature and turn on the power switch. The power lamp is lit when the power heater is on. 3. The temperature scale shows an approximate figure.
5. Maintenance Daily maintenance/Weekly maintenance Do not disjoint the unit Never modify the unit. There are some highvoltage areas inside the unit, which could cause electric shock. Request Yamato Scientific for internal check, adjustment or repair. Maintenance Daily Maintenance: Turn power switch off and disconnect main power cord before attempting any maintenance. Once the water bath has cooled to room temperature you may empty the water.
6. After Sale Service and Warranty Request for Repair When you request repair If any troubles should occur, stop the operation immediately, turn the power off, pull the power cord out and contact Yamato Scientific’s Technical Service Department. Necessary information ● Model Number ● Serial Number ● Date of Purchase ● Distributor Name ● Information on difficulties Warranty ● Keep your warranty card for future references.
7. Replacement Parts Name of Parts Parts Number Specification Fuse 2-10-006-0002 Φ6.4×30 250V 4A (for BM110) Fuse 2-10-006-0003 Φ6.4×30 250V 7A (for BM210) Thermometer 5-02-001-6016 (for BM110/210) Drain Plug 7-32-001-6005 (for BM210) O Ring 4-21-002-6020 P12.
8.