54
Specification
Product Name Coolline
Model CLH301 CLH400 CLH600
Circulation unit Circulation in the external open system
Usable ambient temp.
5℃~30
Temperature control
range
-10℃~80 -15℃~80
Temperature setting
range
-15℃~85 -20℃~85
Temperature
adjustment accuracy
±0.1
Refrigerator
Approx.450W at15 Approx.570W at15 Approx.820W at15
Maximum flow rate
(pump capacity)
10/11 L/min 15/17 L/min
Performance
Maximum head
(pump capacity)
4.9/6.9m 8/11m
Bath SUS304
Temperature control
system
PID control by microcomputer
Sensor
Platinum resistance bulb Pt100Ω
Setting method Digital setting by menu keys and up/down keys
Overheating
prevention sensor
K-thermocouple
Stainless pipe heater
Heater
750W 900W 1500W
Air-cooled sealed type rotary
Refrigerator
200W 350W 600W
Cooling medium HFC R404A 300g HFC R404A 370g HFC R404A 475g
Cooling coil Brass
External circulating
nozzle size
Both discharge and return ports: φ13
Magnet pump
Circulation pump
20W 45W
Configurations
Timer
1min to 99h59min, 100 to 999h50min digital setting,
Quick auto stop, Auto stop, Auto start
Safety devices
Self-diagnostic functions (Memory error, Failure of sensor, Heater disconnection,
Failure of overheating prevention device, SSR short-circuit, At error, Automatic
overheating prevention device), Key lock Function, Earth leakage breaker,
Overheating prevention device, Bypass for circulating pump protection, dustproof
filter for condenser
Other functions
Operation monitor, Drain cock, Temperature output terminal,
RS485communication function
Inner bath
dimensions
(dia.×height)
φ120×200 mm φ150×200 mm φ150×200 mm
External dimensions
(W×D×H)
380×460×500 mm 380×460×720 mm 380×565×720 mm
Bath capacity 1.5L 3L
100V Power supply
(50/60Hz)
12A 15A 25A
Standard
Weight
Approx.40 Approx.45 Approx.60
Optional accessories
1-meter-long insulation hose: 1, 1-meter-long insulation hose with a priming
pump: 1, Drain hose 0.5m: 1, Wire clamp: 4, Instruction manual