5. Behavior after Power Restoration
When having a blackout during operation, the controller resumes the following
operations after the power restoration.
5.1
5.15.1
5.1
When having a blackout during the program operation
When having a blackout during the program operation When having a blackout during the program operation
When having a blackout during the program operation
The controller automatically resume the program operation where it left at the
power shutdown. In case that the temperature inside the chamber is outside the
specified temperature range based on the setpoint temperature, the controller goes
to the FORECED WAIT STATE until the
temperature inside the chamber comes back to the specified temperature range.
When selecting the indication of the remaining time by pushing the Display key in
this condition, the sub display shows
.
The timer built-in the controller does not count a period of having a blackout as
running time.
5.2
5.25.2
5.2
When having a blackout during the Auto-Stop operation
When having a blackout during the Auto-Stop operation When having a blackout during the Auto-Stop operation
When having a blackout during the Auto-Stop operation
The controller automatically resume the Auto-Stop operation where it left at the
power shutdown. In case that the temperature inside the chamber is outside the
specified temperature range based on the setpoint temperature after the power
restoration, the controller goes to the FORECED WAIT STATE until the
temperature inside the chamber comes back to the specified temperature range.
When selecting the indication of the remaining time by pushing the Display key in
this condition, the sub display shows
.(Forced Wait)
In case that the operation stop time is set in a period of time, the timer built in the
controller does not count a period of a blackout as running time. On the contrary,
in case that the operation stop time is set in hours, the timer built in the controller
counts a period of a blackout as running time.
When the operation stop time reaches during a blackout, the controller stops
running just after the power restoration.
5.3
5.35.3
5.3
When having a blackout while the operation is in standby condition
When having a blackout while the operation is in standby condition When having a blackout while the operation is in standby condition
When having a blackout while the operation is in standby condition
In case that the operation start time is set in a period of time, the timer built in the
controller does not count a period of a blackout as standby time. On the contrary,
in case that the operation start time is set in hours, the timer built in the controller
counts a period of a blackout as standby time.
When the operation start time reaches during a blackout, the controller starts
running just after the power restoration.
5.4
5.45.4
5.4
When having a blackout during the Fixed temperature operation and a soak
When having a blackout during the Fixed temperature operation and a soak When having a blackout during the Fixed temperature operation and a soak
When having a blackout during the Fixed temperature operation and a soak
period of the Auto-Start operation
period of the Auto-Start operationperiod of the Auto-Start operation
period of the Auto-Start operation
The controller resumes running toward to the preset temperature after the power
restoration.