User guide

23
Operation Method
Program Operation
This operation is used to change the temperature according to the setting temperature and time.
Program types
Six patterns of program types maximum can be input.
PrG1 - 1 program pattern using 30 steps maximum can be created.
PAt1
PrG2
PAt2
2 program patterns using 15 steps maximum can be created.
PAt1
PAt2
PrG3
PAt3
3 program patterns using 10 steps maximum can be created.
Before inputting program
Input program patterns before program operation.
Check the number of steps in a created program and their setting temperature/time. Use the
program preparation sheet in pages 28and 29 to check.
Check the temperature rise/fall capability of the unit. Set the time within the capability above.
Suppose, for instance, that in the unit which has capability of increasing or decreasing
temperature by 3 within ten minutes, about 35 minutes is needed to increase or decrease
temperature by 10 from current temperature
Repeat function:
Repeat function is useful in case the operation uses the program repeating the same program
steps. Refer to page 27 for the function.
Check if the controller has sufficient free pattern for the number of steps to be created. The steps,
however, using the repeat function mentioned above are not counted.
Temperature fall/rise curve for IJ type
The temperature fall curve and temperature rise curve for IJ type are shown below.
The numeric value indicates the necessary time between temperatures. Temperature stability
time after reaching to the setting temperature is necessary to be added. Make sure to conduct a
test run before setting the optimum time.
Condition: room temperature 20
, no load (unit: minute)
IJ201 IJ300/300W
Fall Rise Fall Rise
60
- 10 - 10
50
15 10 15 10
40
20 10 20 10
30
10 10 10 10
20
15 6 10 10
10
25 5 20 10
0
40 - 45 -
Temp.
Time
Stop Start