User guide

24
Operation Method
Program Operation
Program creation
The program pattern below is explained as an example.
1. Program pattern example
Temp.
40
30
20
10
0
Step 1 2 3 4 5 6 7 8 9
Temp.()
40 40 35 35 25 25 15 10 10
Time(min.)
40 30 15 30 10 30 30
Repeat function
The number of steps is not counted.
15 30
POWER
1. Turn on the power
Turn on the power switch of the unit.
The display on the controller lights on.
The initial screen is displayed for about four seconds, then the
measurement temperature (temperature in furnace) is displayed.
The initial screen displays the software version information, sensor
used and setting temperature of overheating prevention device.
2. Select program mode/program pattern
Press the PROGRAM key once.
The measurement temperature display screen displays the
previous program mode.
Press the PROGRAM key again to display the next program
mode. (PrG1PrG2PrG3)
Select the mode and press the ENTER key.
When PrG1 is selected, the measurement temperature display
screen displays "End".
When PrG2 is selected, the measurement temperature display
screen displays the program pattern "PAt1". For the pattern of
PrG2, select "1" or "2" using the "▲▼". Press the ENTER key
again. The measurement temperature display screen displays
"End".
When PrG3 is selected, the measurement temperature display
screen displays "PAt1". For the pattern of PrG3, select "1", "2"
or "3" using the "▲▼". Press the ENTER key again. The
measurement temperature display screen displays "End".
Any of PrG1. PrG2 or PrG3 can be selectable in the program
example above, where nine steps maximum are used.