User guide

26
Operation Method
Program Operation
5. Start program operation
Press the START/STOP key for about one second. The
program operation previously set starts.
The PROGRAM lamp lights on and the setting temperature
screen displays the step currently under operation.
Press the "" to check the setting temperature and residual time of
step currently under operation on the setting temperature screen.
6. End program operation
Buzzer continues to sound for about five minutes at operation
stop.
The measurement temperature screen displays the character
"END", which indicates the termination of program.
Press the START/STOP key to return to the initial screen.
Timer function:
The maximum setting time is "999 hours and 50 minutes".
The time can be set in increments of a minute under 99 hours
and 59 minutes.
It can be set in increment of ten minutes over 100 hours.
The "▼▲"can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment
is needed.
To correct or check setting…
Press the FIXED TEMP key to correct the created program or to check the setting value. The screen
returns to the former one, where correction or check is possible.
Last screen is displayed when the FIXED TEMP key is once pressed.
Note: Correction or check should be made on the program setting screen.
Wait operation in program operation
The succeeding step does not start in case the measurement temperature does not reach to, or exceeds
the setting temperature when a program goes to the next step in program operation. This unit, however,
is previously set to carry out the next step if the measurement temperature is within ±1 of the setting
temperature.