User guide

34
Maintenance Method
Daily Inspection and Maintenance
For the safety use of this unit, please perform the daily inspection and maintenance without fail.
Using the city water to this unit might attach dirt. Do inspect and maintain this point while
performing daily inspection and maintenance.
WARNING!
Disconnect the power cable from the power source when doing an inspection or maintenance
unless needed.
Perform the daily inspection and maintenance after returning the temperature of this unit to the
normal one.
Do not disassemble this unit.
CAUTION!
Use a well-drained soft cloth to wipe dirt on this
unit. Do not use benzene, thinner or cleanser for
wiping. Do not scrub this unit. Deformation,
deterioration or color change may result in.
Monthly maintenance
Check the earth leakage breaker function.
1. Connect the power cord.
2. Turn the breaker on.
3. Push the red test switch by a ballpoint pen etc.
4. If there is no problem, the earth leakage breaker will be turned off.
Cleaning of air intake filter
Clogging on the filter degrades the performance of the unit. It may also cause the failure of unit.
The degree of clogging on the filter varies depending on the surrounding environment or used
hours. Clean it regularly as necessary.
Refer to the Page 9 for the location of air intake filter.
For any questions, contact the dealer who you purchased this unit from, or the nearest sales
division in our company.