User guide

35
Maintenance Method
Daily Inspection and Maintenance
Cleaning of radiating fin
Unplug the power plug at cleaning
Clogging on the fin degrades the performance of the
unit. It may also cause the failure of unit.
The degree of clogging on the fin varies depending on
the surrounding environment or used hours. Clean it
regularly as necessary.
Loosen the two fixing screws on the ceiling board of unit
and remove the board. For the IJ201 type, remove the
side board.
Disconnect the lead wire of fan and remove the four set
screws from it.
The radiating fin appears after the fan is removed.
Remove the dust on the surface of fin with a vacuum
cleaner.
Assembling should be made after cleaning in the
inverse order above.
CAUTION!
Be careful not to squash the fin at cleaning.
Our company’s service division offers cleanings of radiating fin at cost. Please contact us if
necessary.
If you have any questions, please refer them to your retailer or any of our branch offices.
Set screw
Radiating fin