User Manual

IM 01E20C02-01E
11-3
11. OUTLINE
Accuracy (Combined with AXF Remote Flowtube)
Pulse Output:
PFA/Ceramics Lining:
Size mm
(in.)
V 0.3 (1)
1.0 mm/s
0.5 mm/s
0.5 mm/s
0.3 V 10
(1) (33)
V 0.15 (0.5)
V 0.15 (0.5)
0.15 V 10
(0.5) (33)
0.15 V 10
(0.5) (33)
V 0.15 (0.5)
0.15 V 1
(0.5) (3.3)
0.35% of
Rate
0.5 mm/s
0.35% of
Rate
0.35% of
Rate
2.5 (0.1)
to
15 (0.5)
250 (10)
to
400 (16)
25 (1.0)
to
200 (8.0)
Flow Velocity
V m/s (ft/s)
Flow Velocity
V m/s (ft/s)
Standard
Accuracy
(Calibration
code B)
High grade
Accuracy
(Calibration
code C)
0.18% of Rate
0.2mm/s
1 V 10
(3.3) (33)
0.2% of Rate
T01.EPS
Enhanced dual frequency excitation (Option code HF2):
Standard accuracy 1 mm/ s
Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)
1.2
1.0
0.8
0.6
0.4
0.35
0.2
0
012 6810
V[m/s]
High grade Accuracy
High grade Accuracy
Standard Accuracy
F01.EPS
Pulse
Output
Accuracy
% of Rate
Size 25 mm (1.0 in.) to 400 mm (16 in.)
1.2
1.0
0.8
0.6
0.4
0.35
0.2
0
012 6810
V[m/s]
F02.EPS
Pulse
Output
Accuracy
% of Rate
Standard Accuracy
High grade Accuracy
Polyurethane Rubber/Natural Soft Rubber/EPDM
Rubber Lining
Size mm (in.)
V 0.3 (1.0)
1.0 mm/s
0.3 V 10
(1.0) (33)
0.35% of Rate
25 (1.0) to 400 (16)
Flow Velocity
V m/s (ft/s)
Standard Accuracy
(Calibration code B)
T02.EPS
Enhanced dual frequency excitation (Optional code HF2):
Standard accuracy 1 mm/ s
Size 25 mm (1.0 in.) to 400 mm (16 in.)
1.2
1.0
0.8
0.6
0.4
0.35
0.2
0
012 6810
V[m/s]
High grade Accuracy
High grade Accuracy
Standard Accuracy
F01.EPS
Pulse
Output
Accuracy
% of Rate
Current Output “”: Pulse output accuracy plus 0.05% of Span
Repeatability:
0.1% of Rate (V 1 m/s (3.3 ft/s))
0.05% of Rate 0.5 mm/s (V < 1 m/s (3.3 ft/s))
Maximum Power Consumption:
Combined with AXF Remote Flowtube: 12 W
Insulation Resistance(*1) :
Between power supply terminals and ground
terminal: 100M at 500V DC
Between power supply terminals and input/output/
excitation current terminals : 100M at 500V DC
Between ground terminal and input/output/excitation
current terminals: 20M at 100V DC
Between input/output/excitation current terminals
:20M at 100V DC
Withstand Voltage(*1) :
Between power supply terminals and groundterminal:
1390V AC for 2 seconds
Between power supply terminals and input/output
terminals: 1390V AC for 2 seconds
Between excitation current terminal and ground terminal:
160V AC for 2 seconds
Between excitation current terminal and input /output
terminals: 200V AC for 2 seconds
CAUTION
*1: When performing the Insulation Resistance Test or the
Withstand Voltage Test, please obey the following
caution.
Following the relevant test, wait for more than 10
seconds after the power supply has been turned off
before removing the cover.
Remove all wires from terminals before testing.
When the power terminal has a lighting protector
(optional code A), remove the short bar at the ground
terminal.
After testing, be sure to discharge by using a resis-
tance and return all wires and the short bar to its
correct position.
Screws must be tightened to a torque of 1.18 N-m or more.