7. OPERATION VIA BRAIN TERMINAL (BT200)
(2) Upload/download function of BT200
Upload function is used when the parameters of one
AXFA14 are copied to the BT200. And download
function is used when the parameters copied to the
BT200 are set to another AXFA14.
For details, refer to BT200 User’s Manual (IM
01C00A11-01E).
The targeted parameters for upload and download are
following.
T0705.EPS
Item
C11/B20
C31
C32
C40/B21
C41/B22
C42/B23
C43/B24
Name (BRAIN)
FLOW DAMPING
SIZE UNIT
NOMINAL SIZE
FLOW UNIT
TIME UNIT
FLOW SPAN
FLOW DECIMAL
Item
D10/B30
D11/B31
E10/B32
E11/B33
H10/B40
H11/B41
H12/B42
Name (BRAIN)
TOTAL UNIT
TOTAL SCALE
PULSE UNIT
PULSE SCALE
DISP SELECT1
DISP SELECT2
DISP SELECT3
7.2.3 BT200 Screens & Flow Rate
Data Display
Use the following procedure to display flow rate data
on the BT200.
The display of flow rate data is updated every 5 seconds.
When the BT200 is turned
on, the message “Please
wait ...” is displayed for
several seconds, and then
the screen on the left is
shown.
Press the key.
or
Press the key
to display initial data as
shown on the left.
Press the ( )or
key.
The tag number
specified upon
ordering is found
here.
F0706-01.EPS
–––WELCOME–––
BRAIN Terminal
ID:BT200
Check cable connection
and press the Ent key.
UNTIL FEED
PARAMETER
01:MODEL
AXFA11-A00
02:TAG NO
FI-1101
03:SELF CHECK
GOOD
OK
OK
The Flow Rate Screen
will be displayed.
A maximum of three items
of data can be displayed.
The or
key can be used to move
the cursor up and down.
(Note that the key
can be used to move the
cursor to the top of the
next screen.
Press the
( ) key
to return to Menu Screen
(Major Item Parameter
Search Mode).
F0706-02.EPS
( )
The Menu Screen (Major
Item Parameter Search
Mode) will be displayed.
If the key is
pressed while A: DISPLAY
is selected by the highlight
bar, the Parameter Screen
(Sub-Item Parameter
Search Mode) will be
displayed.
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
PARAMETER
A10:FLOW RATE(
%
)
80.0
%
A20:FLOW RATE
100m3/h
A21:FLW RATE(mA)
16.800mA
DATA DIAG PRNT ESC
ESC
ESC
7.3 Parameter Setting
Using a BT200
This section describes the procedure for setting of
parametes using a BT200.
IMPORTANT
If the power of flowmeter is turned off within 30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
NOTE
Before updating any setting, remember to always
check the data content you want to change as
described in Chapter 6: Parameter Description.