Instruction Manual

<5. EtherLCD>
5-23
IM 11B08A01-01E
F6 (Det #): Sets a detector number.
Pressing F2 (A.Zero) sets the current detector output value to 0 mV exclusively for the A/O
chromatogram. The screen then displays “Auto zero executed.”
Pressing F3 (C.Zero) displays “Zero value clear executed.”
The screen for Oven 1 and Det 1 is displayed as the initial screen.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type None
TCD
FID
FID-MC
FPD
Auto gain None (default)
Individual gain
Whole gain*
* Exclusively for the A/O
chromatogram
Whole gain
value
015
Auto zero Automatic feed (default)
Auto zero value
Disabled
Auto zero
value
–1000.0000 1000.0000 Former chromatogram standard
value
Auto zero
time
0 21600.0 If the value is 0.0 seconds, the
value is set to the zero value
automatically obtained at the
start of the analysis cycle
5.3.2 A/O Operation
Navigating to the A/O Operation screen sets the mode to A/O Manual and all the outputs of 32
A/O channels become 4 mA. On the A/O Operation screen, setting commands are available for
any of the up to 32 A/O channels. Navigating to other screens cancels the A/O Manual mode, and
according to the settings, outputs are made from each A/O channel.
To set the output value of an AO, select the A/O number with the cursor, enter the set value and
press the Set/Ent key. Note that this operation must be done in Manual status and in Stop mode.
11/11/22
15:15:45
A/O Operation
Slot N
AO# Set value
> - 1 A
- 2 A
- 3 A
- 4 A
- 5 A
- 6 A
- 7 A
- 8 A
Menu Slot#
F1 F2 F3 F4 F5 F6
Figure 5.27
Example of the A/O operation screen
F1 (Menu): Displays the Operation Menu screen.
F6 (Slot#): Sets a slot number.
2nd Edition : May 11, 2012-00
5