Instruction Manual
<5. EtherLCD>
5-37
IM 11B08A01-01E
TIP
The Det operation, MC voltage value are refreshed at a one-second interval.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type FID-MC
Det operation 0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
MC voltage
value
0.0 20.0 V
Target MC
voltage
Numeric, period “.” and slash “/”
are enabled: 16 characters
Input format: AA.A/B.B
Output format: AA.A+/–B.B mA
Detector type: FID or FPD
11/11/22
15:15:45
Det Adjustment
11/11/22
15:15:45
Det Adjustment
Oven # 1 Det # 1 ( FID ) Oven # 1 Det # 1 ( FPD )
> - Det operation Off > - Det operation Off
Menu Oven# Det# Menu Oven# Det#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.48
Example of the detector adjustment screen (FID or FPD)
F1 (Menu): Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #): Sets a detector number.
• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Oven # Oven number 1 3
Det # Detector number 1 2
Detector type FID
FPD
Det operation 0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
2nd Edition : May 11, 2012-00