Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
42 of 61
WARNING
Germicidal cloths may cause eye irritation and may be harmful if absorbed through the
skin. Avoid contact with eyes or clothing. Wash thouroughly with soap and water after
handling and before eating, drinking, chewing gum, using tobacco, or using the toilet.
INSPECTION
Once all components have been cleaned and disinfected, they should be visually inspected for any signs
of remaining dirt or soil. If the components appear dirty, repeat cleaning as described above until clean.
Visually inspect for signs of damage, for example cracks or loose connectors in the plastic enclosures.
Do not re-package or re-use components that are showing signs of damage. If components do show
signs of damage, recycle plastic and electronic components in accordance with local regulations.
Additionally, dispose of all other materials in accordance to local regulations.
Ensure all components are completely dry before bagging and re-packaging.
RE-PACKAGING
Power on the HMU, CMU and TTU to test the system and ensure it is functional before packaging.
NOTE
It is important to test whether the system is functional before packaging because
sending a non-functional product to the user will lead to a delay in time before the next
sleep study can be performed and the user could establish a dislike for the product.
Do not repackage components that are not fully functional or otherwise damaged. If components are
not fully functional, please contact the manufacturer for instructions. Additionally, dispose of all other
materials in accordance to local regulations. Repackaging can occur once components have been fully
recharged per instructions in charging the device section.
Place the HMU, CMU and TTU each in a separate small plastic bag. Fold the open end of the bag over
and apply a label marked "clean" to hold the bag closed. This label will be ripped by the user when they
open the bag which will indicate that it has been used.
Place the components back into the reusable case.
UNCONTROLLED COPY - SGS - 30-JAN-19