Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
53 of 61
APPENDIX C ELECTROMAGNETIC COMPLIANCE
The Prodigy Sleep System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Prodigy Sleep System should assure that it is used in such an environment.
When the Prodigy Sleep System is exposed to electromagnetic (EM) disturbances, the system may show
abnormal behaviour. For instance, the measured traces may become obscured, the software may crash,
or, in cases of very high level transient voltage events (e.g. ESD), parts of the system may even become
defective.
Information pertaining to electromagnetic compliance is provided in Table 4 through Table 7
Table 4 - Guidance and manufacturer's declaration - Electromagnetic emissions
Emissions Test
Compliance
Electromagnetic environment
RF Emissions CISPR 11
Group 1
The Prodigy Sleep System uses
RF energy only for its internal
function. Therefore, its RF
emissons are very low and not
likely to cause any interference
in nearby electronic equipment.
RF Emmissions CISPR 11
Class A
The EMISSIONS characteristics
of the Prodigy Sleep System
make it suitable for use in areas
in industrial areas and hospitals
(RF Emmissions CISPR 11 class
A). Harmonic emissions and
Voltage Fluctuations/Flickers
emissions are not applicable for
the device.
Harmonic emissions IEC 61000-
3-2
N/A
Voltage Fluctuations/ Flicker
Emissions IEC 61000-3-3
N/A
Emissions Test
Compliance
Electromagnetic environment
RF Emissions CISPR 11
Group 1
The Prodigy Sleep System uses
RF energy only for its internal
function. Therefore, its RF
emissons are very low and not
likely to cause any interference
in nearby electronic equipment.
RF Emmissions CISPR 11
Class B
The EMISSIONS characteristics
of the Prodigy Sleep System
Harmonic emissions IEC 61000-
N/A
UNCONTROLLED COPY - SGS - 30-JAN-19