Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
55 of 61
typical commercial or
hospital equipment.
Voltage dips, short
interruptions and
volatage variations on
power supply input
lines IEC 61000-4-11
0 % U
T
(100 % dip in U
T
) for 0, 5
cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°.
70 %
U
T
(30 dip in U
T
) for
25 cycles
0 %
U
T
(100% dip in U
T
)
for 5 sec
N/A
Mains power quality
should be that of a
typical commercial or
hospital equipment. If
the user of the Prodigy
Sleep System requires
continued operation
during power mains
interruptions, it is
recommended that the
Prodigy Sleep System
be powered from an
uninterruptable power
supply or a battery.
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be a levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application test level.
The Prodigy Sleep System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Prodigy Sleep System should assure that it is used in such an environment.
When the Prodigy Sleep System is exposed to electromagnetic (EM) disturbances, the system may show
abnormal behavior. For instance, the measured traces may become obscured, the software may crash, or, in
case of very high level transient voltage events (e.g. ESD), parts of the system may even become defective.
Immunity Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6Vrms for ISM bands
N/A
Portable and mobile RF communications
equipment should be used no closer to
any part of the Prodigy Sleep System
including cables, than the recommended
separation distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation distance
= 1.2; from 150 kHz to 80 MHz
UNCONTROLLED COPY - SGS - 30-JAN-19