Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
56 of 61
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 6 GHz
27V/m
385 MHz
28V/m
450 MHz
9 V/m
710/745/780 MHz
28 V/m
810/870/93O MHz
28 V/m
1720/1845/1970 MHz
28 V/m
2450 MHz
9 V/m
5240/5500/5785 MHz
3 V/m
= 1.2; from 80 MHz to 6 GHz
= 2.3P ; from 80 MHz to 6 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
The Prodigy Sleep System is fairly
sensitive to conducted and radiated RF.
Disturbance of the Prodigy Sleep System
trace is possible at and below the
specified test level. It may be necessary to
relocate the Prodigy Sleep System or to
apply shielding.
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
should be less than the
compliance level in each frequency range
b
Interference may occur in the vicinity of
known RF transmitting devices and
equipment marked with the following
symbol:
NOTE 1 - At 80 MHz and 6 GHz, the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Prodigy Sleep System is
used exceeds the applicable RF compliance level above, the Prodigy Sleep System should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the Prodigy Sleep System.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.