Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
24 of 61
ii. Level 2
iii. HMU Only
d. HMU MAC Address
e. CMU MAC Address (if applicable)
f. Oximeter MAC Address (if applicable)
4. Press ‘Save Study Configuration’
5. Press ‘Enter User Mode’
6. Power off TTU
Configure Sleep Study via Remote Server
A technician can configure a sleep study via remote server. If the technician selects this method, follow
the steps below:
1. Power on TTU, Prodigy Sleep System App will launch automatically
2. Press ‘Run TTU Diagnostics’
a. Should get message that states ‘This tablet is suitable for use’
b. If you get a message that states ‘This tablet is not suitable for use’ refer to Section 6
Troubleshooting to resolve the issue
3. Press ‘Configure Server URL’ and enter the URL of the Remote Server
a. This step only needs to be completed the first time a sleep study is configured via the
remote server
4. Press ‘Fetch Configuration’
5. Press ‘Enter User Mode’
6. Power off TTU
Upload Sleep Study to Remote Server
Uploading a sleep study to a remote server is only available if the study was configured via remote
server. If a user is unable to upload the study to the remote server or if the upload isn’t fully complete,
the technician should:
1. Power cycle the TTU if the tablet is returned powered on
2. In technician mode, press ‘Process Upload Queue’
3. Study will upload to remote server and the sleep study data will automatically be removed from
the TTU
a. If the TTU is unable to upload the study refer to Section 6 Troubleshooting to resolve
the issue
Manually Download Sleep Study
To manually download a sleep study the technician should:
1. Power on TTU, Prodigy Sleep System App will launch automatically
UNCONTROLLED COPY - SGS - 30-JAN-19