Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
44 of 61
EXPECTED SERVICE LIFE
The HMU and CMU devices are powered by a rechargeable polymer lithium-ion battery with an 800mAh
capacity. The battery will self-discharge over time, and if unused for an extended amount of time at low
charge the battery could be irreparably damaged. Assuming battery is full charged prior to being placed
in long term storage, and an average discharge rate of 4% per month, the minimum expected shelf life
of the battery is 25 months.
The battery manufacturer recommends batteries to be charged to 50-70% prior to being placed in long
term storage and undergo a full charge/discharge cycle every 3 months. Following the manufacturer
recommendations will extend the shelf life of the battery.
The expected service life of all accessories is included in
Table 1.
Table 1 – Expected Service Life of Accessories
Accessory
Expected Service Life
Effort Belts with Dual Connector
2 years
Nasal Cannula
1 use
Dual Snap Lead Cables for Legs
2 year
Dual Snap Sensor Pads for Legs
1 use
Single Snap Lead Cables for Chest
2 year
Single Snap Sensor Pads for Chest
1 use
USB Wall Adapter with USB-C Cable
2 year
Dual Snap Sensor Pad
1 use
Single Snap Sensor Pads
1 use
Wired Sensor Pads
1 use
USB Wall Adapter with Micro USB Cable
2 year
UNCONTROLLED COPY - SGS - 30-JAN-19