Operator's Manual

Table Of Contents
Operator Manual Prodigy Sleep System
Revision: 04; CCRF-625
Effective Date: 30-Jan-19
Document Number: YST-UDoc-PR2-001-04
46 of 61
o If device still does not connect power cycle TTU and all peripherals
o If device still does not connect ensure device is assigned to TTU
Oximeter displays Device Error on the TTU
o Power cycle device
o If device still does not connect power cycle TTU and all peripherals
HMU signal(s) do not transition to 'Good'
o Ensure all electrodes connected to HMU
o Ensure all electrodes securely attached to skin
o Remove and replace electrodes according to Section 4
CMU EMG or EKG signal(s) do not transition to 'Good'
o Ensure all electrodes connected to CMU
o Ensure all electrodes securely attached to skin
o Remove and replace electrodes according to Section 4
CMU Belt signal(s) do not transition to 'Good'
o Ensure belt(s) connected to CMU
o Ensure belt is setup according to Section 4
HMU signals good but still shows 'Needs Attention'
o If wireless indicator is less than full move the TTU closer to the user
o If battery indicator is less than 95% HMU will need to be charged prior to running study
CMU signals good but still shows 'Needs Attention'
o If wireless indicator is less than full move the TTU closer to the user
o If battery indicator is less than 95% CMU will need to be charged prior to running study
Oximeter signals good but still shows 'Needs Attention'
o Replace batteries in Oximeter
Peripherals good but 'Start Study' disabled
o Ensure the TTU is connected to charger
Signal goes bad during the night
o If belt, cannula, or oximeter signal report as bad, adjust according to Section 4
o If EEG/EOG/EMG/EKG signal report as bad, ensure the electrode is connected to device
(touchproof, snap)
o If signal continues to report as bad, allow study to continue with bad signal