Although new parameter values are stored  immediately  when  the 
command is issued, they do not take effect until the sensor enters 
OPERATE MODE.
  COMMAND   DESCRIPTION
  SET01nn Set OUTPUT MODE
    01    Enable voltage output
    02    Enable RS-232
    04    Enable RS-485 full duplex
    08    Enable RS-485 half duplex
    16    Enable RS-485 output only
  SET02nn Set OUTPUT FORMAT
    01    RMYT (default)
    02    ASCII
    03    ASCII POLLED
    04    NMEA
    05    SDI-12 CONTINUOUS
    06    SDI-12 STANDBY
  SET03nn Set BAUD RATE
    12    1200
    48    4800
    96    9600
    38    38400
  SET04nn Set ASCII WS UNITS
    01    MPH 
    02    KNOTS
    03    KMPH
    04    M/S
  SET05c    ASCII character sensor address (0-9, A-Z, a-z)
  SET06nnn  Wind speed threshold for polar output (cm/s)
  SET07nnnnn  Wind speed scale (nnnnn/10000)
  SET08nnnnn  Direction offset (±nnnnn degrees x 10)
  SET09nn  Damping factor
  SET10nnn  Output delay (0-999 x 8.192 milliseconds)
  SET11nn  Direction VOUT (36=0-360, 54=0-540 degrees)
  SET12nn  Sample count (00=auto or 03-31)
  SET13nn  ASCII and Vout format (00=polar, 01=UV)
  SET14nn  Vout Status Code (1=none, 2=0V, 3=FS)
  CALnn    Force Vout (00=0V, 01=5V)
  XX      Go to OPERATE MODE
  RPT      Report parameter settings
  ??      Help list
6.4 COMMAND DETAILS
6.4.1 SET01nn SET OUTPUT MODE
This parameter independently enables and disables voltage output, 
RS-232, and RS-485 connections. For example, to enable voltage 
output, RS-232, and RS-485 full duplex, add the code for each one: 
01 + 02 + 04 = 07, SET0107. To enable only RS-232: SET0102. If 
RS-485 is used, choose only one RS-485 option (04, 08, or 16).  See 
section 6.3 for codes.
To reduce power consumption, enable only those outputs that are 
needed. When SDI-12 format is  selected with the SET02nn com-
mand, voltage output, RS-232, and RS-485 connections are disabled 
and SDI-12 is enabled.
6.4.2 SET02nn SET OUTPUT FORMAT
This parameter determines the serial output data format. See section 
5.0 for operational details. When selecting some formats, parameters 
are automatically changed as listed below:
  FORMAT  ADDITIONAL
  CODE  PARAMETERS CHANGED
  01  RMYT  BAUD = 9600
      OUTPUT MODE = 07
      THRESHOLD = 25 cm/s
      OUTPUT DELAY = 01
      SAMPLE COUNT = 31
  02  ASCII
  03  ASCII POLLED  None
  04  NMEA  BAUD = 4800
      OUTPUT MODE = 19
      THRESHOLD = 25 cm/s
      OUTPUT DELAY = 122
      SAMPLE COUNT = 31
  05  SDI-12 CONT
  06  SDI-12 STANDBY  1200 baud
Both SDI-12 settings automatically disable voltage output, RS-232, 
and  RS-485  connections.  With  either  POLLED  ASCII  or  SDI-12 
STANDBY, set the Damping Factor to 0 (see section 6.4.9).
6.4.3 SET03nn SET BAUD RATE
All RS-232 and RS-485 communication use the baud rate set with 
this parameter. Power consumption is affected by this setting since 
it inuences communication duration.
6.4.4 SET04nn SET ASCII WIND SPEED UNITS
Sets wind speed units for ASCII, ASCII POLLED, and SDI-12 polar  
outputs only.
6.4.5 SET05c SET POLL ADDRESS CHARACTER
Sets the sensor address for ASCII POLLED and SDI-12 formats. The 
default is '0' (ASCII 48).
6.4.6 SET06nn SET WIND SPEED THRESHOLD
Sets the wind speed threshold for polar outputs (wind speed and 
direction) to suppress scattered wind direction indications  in  calm 
air. Winds below the threshold are reported as zero, wind direction 
is  held  at  last  value  when  wind  speed  was  above  threshold.  Set 
threshold in centimeters per second (m/s x 100). The default setting 
is 25 cm/s (0.25 m/s, 0.56 mph).
6.4.7 SET07nnnnn SET WIND SPEED MULTIPLIER
All wind speed measurements are multiplied by this parameter. The 
default value is 10000 for a multiplier of 1.0000. This may be used to 
individually calibrate each sensor for improved accuracy.
6.4.8 SET08nnnnn SET WIND DIRECTION OFFSET
Use this parameter to add or subtract a wind direction offset. Value 
is degrees x 10 and may be positive or negative. Wind direction is 
always re-scaled to a 0-360 range after offset is applied. The default 
value is 00000.
6.4.9 SET09nn SET DAMPING FACTOR
Wind measurement outputs are damped using the following formula:
  S
damped 
= [(d-1) * S
damped
 + S
sample
] / d
  where:
  S
damped 
= New or last damped wind speed
  S
sample
  = New wind speed speed
  d  = Damping factor