Owner's manual
 Page 6
85000-90
The  default  value  is  4.  When  set  to  00,  no  damping  is  applied. 
Higher damping factors slow the rate at which the sensor indicates 
wind changes.
6.4.10 SET10nnn SET OUTPUT DELAY
Sets the amount of time between measurements in increments of 
8.192 milliseconds. The default value is 122 ( 0.9994 seconds). Lower 
values (less delay) increase power consumption when continuous 
measurements are taken.
6.4.11 SET11nn SET DIRECTION VOUT
Sets wind direction voltage output scale to 0-360 or 0-540 degrees.  
Use the 0-540 scale if possible. When set to 0-360, wind directions 
uctuating about north cause voltage output swings between zero 
and full scale. Data loggers or display systems may sample during 
these transitions causing erroneous readings. A 0-540 scale elimi-
nates output transitions with northerly wind.
Data in 0-540 form may be re-scaled to 0-360  by subtracting 360 
degrees from any value greater than or equal to 360. The default 
parameter  setting  is  0-360  for  systems  that  cannot  re-scale  the 
0-540 output.
6.4.12 SET12nn SET SAMPLE COUNT
All measurements are calculated from the median of 3 to 31 sample  
sets.  More samples consume more power. More samples are 
required  at  higher  wind  speeds  or  in  conditions  where  adequate 
samples  are  difcult  to  acquire.  If  SAMPLE COUNT is  set  to  00, 
the sensor automatically uses the minimum number of sample sets 
needed to produce valid data.  However, when powered up for the 
rst time or when polling at intervals longer than a few seconds, the 
sensor may require more than one measurement cycle to determine 
the optimum number of sample sets.
In applications where power consumption is not an issue or where 
the sensor must acquire results on the rst measurement cycle, the 
SAMPLE COUNT may  be set manually to the highest  anticipated 
wind speed (m/s) up to 31 which is the maximum SAMPLE COUNT 
value. Wind speeds above 31 m/s will use 31 sample sets.
6.4.13 SET13nn SET ASCII and Vout FORMAT
Serial ASCII and ASCII POLLED formats may be set to provide wind 
data in either Polar (speed and direction) or Cartesian (UV) form. This 
setting also determines the whether voltage outputs are in Polar or 
Cartesian format.
00= Polar, 01=Cartesian (UV).
6.4.14 SET14n SET VOLTAGE OUTPUT STATUS 
Serial ASCII and ASCII POLLED formats report a STATUS CODE 
where non-zero values indicate insufcient samples or measurement 
error. SET14n determines how the STATUS CODE is handled in the 
voltage output.
1 = None (ignore STATUS CODE)
2 = Set Vouts to 0V when STATUS is non-zero
3 = Set Vouts to 5V full scale when STATUS is non-zero
6.4.15 CALnn FORCE VOUT
This command forces both voltage outputs channel to either 0mV 
or 5000 mV. This may be used to calibrate or check the operation of 
external data acquisition devices. 00=0 mV, 01=5000 mV
6.4.16 XX, RPT, and ??
XX  Returns the sensor to OPERATE MODE.
RPT  Reports the latest parameter settings.
??  Prints a list of commands.
7.0 EXAMPLE SETTINGS
Use the 2DSETUP program for the following suggested settings. Not 
all possible setting combinations are shown. 
 When heaters are enabled the OUTPUT DELAY is forced to 0 and 
SAMPLE COUNT is forced to 31. 
7.1  VOLTAGE OUTPUT ONLY, MINIMUM POWER
  Serial Output Format: ASCII
  RS-232  Disabled
  RS-485  Disabled
  Voltage Output:  Enabled
  Output Delay:  122
  Sample Count:  0
7.2  RS-232 ONLY, CONT ASCII OUTPUT, MINIMUM POWER
  Serial Output Format: ASCII
  RS-232  Enabled
  RS-485  Disabled
  Voltage Output:  Disabled
  Output Delay:  122
  Sample Count:  0
7.3  RS-232 ONLY, POLLED ASCII OUTPUT, MINIMUM POWER
  Serial Output Format: ASCII POLLED
  RS-232  Enabled
  RS-485  Disabled
  Voltage Output:  Disabled
  Sample Count:  0
7.4  RS-232 ONLY, POLLED ASCII OUTPUT
  Serial Output Format: ASCII POLLED
  RS-232  Enabled
  RS-485  Disabled
  Voltage Output:  Disabled
  Sample Count:  Set to 0 if polling frequently. Set to 
    highest anticipated wind speed (m/s) up 
    to 31 if polling infrequently. Average
    current consumption is a function of
    polling rate and sample count.
7.5  SDI-12 CONTINUOUS
  Serial Output Format: SDI-12 CONT
  Output Delay:  0
  Sample Count:  0
7.6  SDI-12 STANDBY
  Serial Output Format: SDI-12 STANDBY
  Output Delay:  0
  Damping Factor:  0
  Sample Count:  Set to 0 if polling frequently. Set to 
    highest anticipated wind speed (m/s) up 
    to 31 if polling infrequently. Average
    current consumption is a function of
    polling rate and sample count.