Owner's manual

Calibrate Section 6
YSI Incorporated YSI 556 MPS
Page 35
Sensor to Calibrate Upright Upside Down
Conductivity 55ml 55ml
pH/ORP 30ml 60ml
Table 6.1 Calibration Volumes
6.2 Calibration Procedures
6.2.1 Accessing the Calibrate Screen
1. Press the On/off key to display the run screen.
2. Press the Escape key to display the main menu screen.
3. Use the arrow keys to highlight the Calibrate selection.
Figure 6.1 Main Menu
4. Press the Enter key. The Calibrate screen will be displayed.