Instruction Manual

System 182 PROFIBUS connection
7 - 11
ba76024e01 02/2012
the status of the main measured value, bits 3-0 encode the status of the
secondary measured value.
Other data All other data always consist of only a single byte (Int8).
7.6 PROFIBUS error diagnosis
A simple diagnosis of the PROFIBUS communication is possible via
the measured value and status display (status online/offline, see
section 4.2).
7.7 PROFIBUS error elimination
Note
Here you will find causes and actions to take of errors concerning the
PROFIBUS communication only. General errors of the IQ S
ENSOR NET
system are dealt with in chapter 9 W
HAT TO DO IF ....
Data transmission
between the Profibus
master and
DIQ/S 182-PR is faulty
The PLC does not
receive any plausible
input data
Cause Remedy
Incorrect wiring of the
PROFIBUS cable on the
connector
Check/change the connections
(see section 7.2)
Use checklist according to
section 7.1
Incorrect setting of the
address
Check/change the setting of
the address (see section 7.3)
Incorrect protocol Check the version of the GSD
file
Adapt the protocol
DIQ/S 182-PR defective Send DIQ/S 182-PR to YSI
Cause Remedy
Input data and output data
are not consistent
When programming the PLC,
define input data and output
data as consistent over the
entire data length
The data interpretation of the
PLC is not correct
Observe the data alignment of
the PLC data representation. If
necessary, exchange the high-
order and low-order bytes word
by word