3.10.2 DIQ/S 182 XT-4-PR/24V (24 V version)
Warning
If the 24 V AC/DC supply is incorrectly connected, it may
represent a danger to life from electric shock. Pay attention to the
following points during installation:
The DIQ/S 182 XT-4-PR must only be connected by a qualified
electrician.
The 24 V AC/DC supply must meet the specifications quoted on
the name plate and in chapter 10 T
ECHNICAL DATA (protective
low voltage SELV).
The DIQ/S 182 XT-4-PR must only be connected in a voltage-
free condition.
A switch or circuit breaker must be provided in the building
installation as an interrupt facility for the DIQ/S 182 XT-4-PR.
The interrupt facility must be
– installed in the vicinity of the DIQ/S 182 XT-4-PR, easily
accessible by the user and
– marked as an interrupt facility for the DIQ/S 182 XT-4-PR.
Note
Rechargeable battery systems should have a deep discharge
protection. The DIQ/S 182 XT-4-PR/24V does not have any built-in
deep discharge protection.
Materials required Wire end sleeves, suitable for the 24 V AC/DC feed line, with
suitable crimping tool
1 x cable gland, matched to the cable diameter (see section 3.9 on
page 3-33).
Tools Cable stripping knife
Wire stripper
Phillips screwdriver
Small screwdriver.
Preparing the 24 V AC/
DC line
1 Cut off the cable to the required length.
2 Strip the cable insulation for approx. 45 mm.
3 Bare the wires 1 and 2 and fit them with wire end sleeves.