Owner's manual
Operation System 182 XT-4
4 - 16
ba76027e01 02/2012
Putting the sensor into
operation after
calibrating
Note
If calibration was not successful, this is indicated by "----" on the display
and a corresponding message with remedial actions appears in the log
book. Follow the instructions and repeat calibration.
5 Confirm the calibration data with g. Calibration is completed
with this. The following display message describes the further
steps to put the sensor into operation again.
6 Confirm with g. The display returns to the measured value
display (the measured value flashes as the sensor is still in the
maintenance condition).
7 If calibration was successful, immerse the sensor in the test
sample.
8 Wait for a stable measured value.
9 Switch off the maintenance condition (highlight sensor and
press
g three times). The measured value has stopped
flashing.