OPERATIONS MANUAL ba76137e01 4110 / 4110-3 PH ELECTRODES WITH GEL ELECTROLYTE 03/2013
4110 For the most recent version of the manual, please visit www.ysi.com. Contact Copyright 2 YSI 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com © 2013 Xylem Inc.
4110 General information General information Automatic sensor recognition The sensor electronics with the stored sensor data is in the connecting head of the electrode. The data include, among other things, the sensor type and series number. With each calibration, the calibration data is written in the sensor and the calibration history is recorded. The data is recalled by the meter when the sensor is connected and is used for measurement and for measured value documentation.
Commissioning, measuring, calibration 4110 Shaft dimensions, material, electrical connection Model Shaft Electrical connection Length [mm] Ø [mm] Material Combination Meter electrode connection connection 4110 120 12 PPE/PS Fixed cable Digital plug 1.5 m 4110-3 120 12 PPE/PS Fixed cable Digital plug 3m Cable length Connection cable Diameter 4.
4110 Commissioning, measuring, calibration – Between measurements, rinse the electrode with the solution that follows. Alternatively, you can also rinse the electrode with deionized water and then carefully dab it dry. To measure in aqueous solutions, it is recommended to immerse the electrode in a vertical or slightly tilted position. Observe the correct depth of immersion and make sure the contact between the junction and test sample is thorough.
Aging 4110 Storage During short measuring breaks Immerse the electrode in reference electrolyte (KCl 3 mol/l, Ag+ free). Prior to the next measurement, shortly rinse the electrode with the test sample or deionized water. Note Do not scratch the pH membrane. Overnight or longer Put the clean electrode in the watering cap that is filled with reference electrolyte (KCl 3 mol/l, Ag+ free). Note pH electrodes must not be stored dry or in deionized water. The electrode could be permanently damaged by this.
4110 Maintenance and cleaning Maintenance and cleaning Cleaning Remove water-soluble contamination by rinsing with deionized water.
Wear parts and accessories 8 4110 ba76137e01 03/2013
4110 Contact Information Contact Information Ordering & Technical Support Telephone: (800) 897-4151 (937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET Fax: (937) 767-1058 Email: environmental@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA Internet: www.ysi.
Contact Information 10 4110 ba76137d01 03/2013
Xylem |'zīləm| 1) The tissue in plants that brings water upward from the roots; 2) a leading global water technology company. We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work.