4110 General information
3
ba76137e01 03/2013
General information
Automatic sensor
recognition
The sensor electronics with the stored sensor data is in the connecting head
of the electrode. The data include, among other things, the sensor type and
series number. With each calibration, the calibration data is written in the
sensor and the calibration history is recorded. The data is recalled by the
meter when the sensor is connected and is used for measurement and for
measured value documentation. Storing the calibration data in the sensor
ensures that the correct slope and asymmetry are automatically used if the
sensor is operated with several meters.
The digital transmission technique guarantees the failure-free communi-
cation with the meter even with long connection cables. If the sensor firmware
is enhanced by YSI, it can be updated via the meter.
Technical Data
General data
Measurement and application characteristics
Model Reference
elektrolyte
Junction NTC Special fatures
4110 Gel Fiber Yes Plastic shaft
4110-3 Gel Fiber Yes Plastic shaft
Model pH measuring
range
Allowed
temperature
range
Membrane
resistance at
25 °C
Typical
application
4110 0.000 ... 14.000 0 ... 80 °C < 1 GOhm Field