Instruction Manual
Table Of Contents
- Warranty
- Getting started
- Buffer Set Selection
- pH Calibration
- pH Measurements
- Temperature Measurements
- mV Measurements
- Troubleshooting
- Specifications
- Accessories/Spare Parts
- Recycling
- Contact Information
- GarantÍa
- CÓmo Empezar
- SelecciÓn Del Juego De Soluciones Buffer
- CalibraciÓn de pH
- Mediciones De pH
- Mediciones De Temperatura
- Mediciones De mV
- SoluciÓn de Problemas
- Especificaciones
- Lista De Piezas De Recambio Recomendadas
- Reciclaje
- InformaciÓn De Contacto
- GARANTIA
- INTRODUÇÃO
- SELECÇÃO DO CONJUNTO DE TAMPÕES
- CALIBRAÇÃO DE pH
- MEDIÇÕES DE pH
- MEDIÇÕES DE TEMPERATURA
- MEDIÇÕES mV
- ResoluÇÃo de Problemas
- ESPECIFICAÇÕES
- LISTA DE PEÇAS SOBRESSALENTES RECOMENDADAS
- Reciclagem
- Eliminação de pilhas
- Informação de contacto

13
14
ACCESSORIES/SPARE PARTS
Item Description
601201 pH Electrode, Lab, 1M
601202 Temperature Probe, Stainless Steel
601203 pH/ATC Temp Combination Probe, Lab, 1M
605170 Electrode stand
603824 pH buffer, assorted case of US buffers. 2 pints ea of
buffer 4, 7 and 10
RECYCLING
YSI is committed to reducing the environmental footprint in the course
of doing business. Even though materials reduction is the ultimate
goal, we know there must be a concerted effort to responsibly deal
with materials after they’ve served a long, productive life-cycle. YSI’s
recycling program ensures that old equipment is processed in an
environmentally friendly way, reducing the amount of materials going
to landfills.
• Printed Circuit Boards are sent to facilities that process and reclaim
as much material for recycling as possible.
• Plastics enter a material recycling process and are not incinerated
or sent to landfills.
• Batteries are removed and sent to battery recyclers for dedicated
metals.
When the time comes for you to recycle, follow the easy steps outlined
at ysi.com.
BATTERY DISPOSAL
The 1200 is powered by alkaline batteries which the user must remove
and dispose of when the batteries no longer power the instrument.
Disposal requirements vary by country and region, and users are
expected to understand and follow the battery disposal requirements
for their specific locale.
SPECIFICATIONS
Display Range Accuracy Resolution
pH 0.00 to 14.00 +/- 0.01 (instrument
only)
0.01 unit
Temperature 0.0 to 100°C +/- 0.2°C 0.1°C
mV -1999 to 1999
mV
+/- 0.05% +/- 1 digit 0.1 mV
Operating Temperature
Range
0 to 50°C (32 to 122°F)
Operating Relative
Humidity Range
up to 85%
Water resistance IP-65 water resistant, splash resistant
Size 150 mm x 203 mm x 72 mm
5.9 in. x 8 in. x 2.8 in.
Weight with batteries 504 grams (1.1 lb)
Input impedance
>10
13
Power Four 1.5 v C-size alkaline batteries
or 115/230V AC adapter
pH temp compensation Auto/Manual 0.0 to 100.0°C
pH buffer recognition Automatic recognition
US (4.01, 7.00, 10.01) or
NIST (4.00, 6.86, 9.18)
pH offset recognition +/- 1.50 pH
pH slope recognition 30%
Connector (pH/temp) BNC/8 pin DIN
Autolock feature Yes
Backlit Display Yes