TruLab pH 1110 What to do if...
ba76155e01 03/2014 39
10.1.3 Error message OFL, UFL
The measured value is outside the measuring range.
The measured value is obviously incorrect.
10.2 ORP
10.2.1 No stable measured value
Cause Remedy
The used buffer solutions do not
agree with the set buffer set
Set different buffer set
or
Use different buffer solutions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions
Cause Remedy
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
Air bubble in front of the junction Remove the air bubble
Air in the junction Extract air or moisten the junction
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Junction contaminated Clean the junction
Pt-ORP electrode contaminated Clean the Pt-ORP electrode