Specifications

Care, maintenance and trouble-shooting
Carl Zeiss Troubleshooting Axiostar plus
4-4 B 40-815 e 12/01
1 Halogen lamp
2Cover
Fig. 4-2 Changing the 6 V, 20 W halogen lamp
When changing the lamp, it is not necessary to
remove the microscope from the ergonomic
base plate (in the event the base plate is being
used).
4.2.2 Changing the 6 V, 20 W
halogen lamp
Do not touch the lamp bulb with your bare
hands; if required, clean the bulb with clean
alcohol before switching it on for the first time
to prevent dirt from burning in.
The following procedure is required to
exchange the lamp:
Switch off lamp supply via on/off switch
and allow lamp to cool down for approx.
15 minutes.
Disconnect the earth-contact plug from the
line and remove the flat plug from the
instrument connector.
Place the disconnected instrument on its
back to make the underside of the pyramid
stand accessible.
Fold down cover (4-2/2) and pull out
defective halogen lamp (4-2/1).
Use the protective cover to hold the new
6 V, 20 W halogen lamp and insert both
lamp pins carefully into the receptacles.
The Axiostar plus microscope is supplied
with factory-aligned illumination. In general,
the illumination need not be adjusted even
when the lamp is exchanged by the
customer.
Fold up cover again, return the stand to the
upright position and reconnect the stand to
the line.