Operating instructions

MC 80
DX
Carl Zeiss
B 40-036 e 03/99 2-13
2.3.3
AUTO mode (Automatic exposure control)
Pressing the <AUTO> button activates automatic
exposure control. The exposure time shown by the
LCD display is determined by the existing light
intensity, the film speed, the method-specific
exposure correction, and the camera factor.
During the exposure, the time indication is running
back to zero.
2.3.3.1
Exposure correction in AUTO mode
Use the <UP> ad <DOWN> buttons to enter the
desired exposure correction for the microscopy
method used, in whole increments. For easier
input, the LCD display indicates the methods and
the correction values assigned to them.
2.3.3.2
Common settings for method-
specific corrections
For brightfield (Dark features on a bright background) BF +1
For phase or interference contrast PH/DIC 0
For darkfield/fluorescence (Bright features on a dark background) DF/FL -2
Depending on the type and size of the features seen before the background it may be necessary to
deviate from the above settings in favor of these:
For brightfield
Few, or very small, dark features on large expanse of bright background BF +2
or
For darkfield / fluorescence
Few, or very small, features on large expanse of dark background DF/FL -3
AUTO
AUTO
EXP.ADJ.
UPDOWN