Technical data
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Operation 111
Use the "∇" and "∆" keys to change the settings in steps.
Each of the three adjustment speeds is variable in a range from 1 (min-
imum) to 10 (maximum) in steps of 0.5.
With low zoom values, optical systems have a large depth of field and the
focussing system has to cover a large adjustment range until the image is
sharply defined. The X-Y coupling needs to be adjusted within a wide
range until the position required is reached. This takes a certain time, and
a high adjustment speed is therefore of advantage in this case.
With high zoom values, on the other hand, optical systems have a small
depth of field. The focussing system has to be precisely positioned to ob-
tain a sharply defined image, and the X-Y coupling needs to be precisely
adjusted within a narrow range to reach the position required. A low ad-
justment speed is therefore preferable here.
Select the focusing speed which suits your specific work method.
Saving parameter settings
As soon as you have entered a parameter setting, it is saved under the
current user ID.
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
Path: To return to the basic mode, press the "MODE" key in the speed
mode or do not press any key for 20 seconds.
To switch to configuration mode 1, simultaneously press the
"MODE" and "STORE" keys.