Technical data

114 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
What to do in an emergency
Failure of a halogen lamp
Caution:
Do not cover the ventilation grid (2)! Make sure that drapes do not cover
the grid. This can lead to overheating of the lamp modules and to lamp
failure.
Note:
Each of the two lamp housings contains a backup lamp which will be au-
tomatically activated if the first lamp fails. The open flap (3) and the yellow
indicator lamp (7) indicate that the backup lamp is operative.
Manual switchover to the backup lamp
Press the button (4) to manually activate the backup lamp.
In the event of failure of the backup lamp:
Warning!
If you change the lamp shortly after it has gone out, the lamp will still be
very hot. Wear heat-protection gloves to avoid burns!
Turn off the ceiling mount (floor stand) at the power switch.
Press button (5), the lamp module (1) is slightly ejected. Remove the
lamp module and change the lamp, or insert the lamp module of the
second illumination system.
Switch the ceiling mount (floor stand) back on. Adjust the brightness
of the illumination on the display field (6) as required.