Technical data

116 Operation
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Failure of the xenon lamp
Caution:
Do not cover the ventilation grid! Make sure that drapes do not cover the
grid. This can lead to overheating of the lamp module and to lamp failure.
Note:
Yellow indicator lamp (4) lights when the lamp has failed. It goes out again
after activation and ignition of the backup lamp.
Activating the backup lamp
Turn off the stand at the power switch before activating the backup
lamp.
The lamp module contains two xenon lamps. The second lamp is used as
a backup lamp which is swung into the illumination beam path when the
first lamp fails.
If the first xenon lamp fails, you can open lamp module (2) as follows:
Press button (3). The lamp module is slightly ejected.
Pull out the lamp module all the way.
Turn knob (1) through 180° until it snaps in. This swings the second
xenon lamp (backup lamp) into the beam path.
Push the lamp module all the way back into the lamp housing.
Turn the stand back on at the power switch.