Failure of the xenon lamp
Caution:
Do not cover the ventilation grid! Make sure that drapes do not cover the
grid. This can lead to overheating of the lamp module and to lamp failure.
Note:
Yellow indicator lamp (4) lights when the lamp has failed. It goes out again
after activation and ignition of the backup lamp.
Activating the backup lamp
• Turn off the stand at the power switch before activating the backup
lamp.
The lamp module contains two xenon lamps. The second lamp is used as
a backup lamp which is swung into the illumination beam path when the
first lamp fails.
If the first xenon lamp fails, you can open lamp module (2) as follows:
• Press button (3). The lamp module is slightly ejected.
• Pull out the lamp module all the way.
• Turn knob (1) through 180° until it snaps in. This swings the second
xenon lamp (backup lamp) into the beam path.
• Push the lamp module all the way back into the lamp housing.
• Turn the stand back on at the power switch.