Technical data
126 Maintenance / Further information
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Changing the halogen lamp
Warning!
When you change the lamp shortly after it has gone out, the lamp will still
be very hot. Wear heat-protection gloves to avoid burns!
Note:
If both halogen lamps (4 and 5) fail during surgery, insert the lamp module
of the second illumination system. For this reason, always make sure be-
fore surgery, that the halogen lamps of the second illumination system are
intact.
To change the lamp, proceed as follows:
• Switch the illumination off using the knob (3). Press the power switch
to turn off the system.
• Press button (2), the lamp module is slightly ejected. Remove the
lamp module (1).
• Remove the defective halogen lamp from its spring-loaded mount.
• Pull the ceramic base (8) from the contact pins of the halogen lamp.