Technical data
128 Maintenance / Further information
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
• Plug the ceramic base (8) onto the contact pins of the new halogen
lamp.
• Insert the new halogen lamp. Make sure you do not touch the lamp
bulb (7) or the interior of the reflector (6).
• Press the halogen lamp into the spring-loaded mount.
• Push the lamp module including the new halogen lamp back into the
unit.
• Press the power switch to turn on the system, and switch on the illu-
mination with the relevant knobs.
Note:
– Only use 12 V, 100 W halogen lamps available under Cat. No.: 38 00
79- 9040
To use the service life of your halogen lamps as economically as possible,
we recommend that you proceed as follows: If the main lamp (4) has
failed, remove it and replace it by the backup lamp (5). Install the new hal-
ogen lamp instead of the backup lamp.