Technical data

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Maintenance / Further information 137
Electrical data
Power connection Only connect the stand to power outlets which are
equipped with a properly connected protective
earth conductor.
Rated voltage 115 VAC (100...125 V
~ ± 10%)
230 VAC (220...240 V
~ ± 10%)
Power consumption 115 VAC max.10 A
230 VAC
max.8 A
Rated frequency 50...60 Hz
Fuses Automatic circuit breaker
Electrical
outlets
power outlet 115/230 VAC , max 5 A
power outlet 115/230 VAC , max 2 A,
(via power switch),
XY coupling
surgical microscope
Electrical
standard
complying with IEC 601-1/EN 60 601-1/ UL 2601-
1; CSA C22.2 No 601.1
Protection class I,
type B equipment
EMC
requirements
complying with EN 60601-1-2: 1993
Class A (hospital)