Technical data
18 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Halogen illumination system
1 Activating the backup lamp
Each of the two lamp housings contains a backup lamp which will be
automatically activated if the first lamp fails. If this automatic function
fails, you can switch on the backup lamp by pressing this button.
2 GG 475 retina protection filter
When operating on the eye, always use a GG 475 protection filter to
ensure that the patient's retina is not exposed to unnecessary (blue)
radiation (risk of retinal injury). The filter knobs have four positions:
3 Yellow indicator lamp
– Lights when the main lamp has failed. In addition, the open flap on
the lamp module indicates that the main lamp has failed. The back-
up lamp is on.
– Blinks when the backup lamp has failed.
4 Manual function
When the manual function has been activated, all electrical control
systems are inoperative. The lamp brightness is automatically ad-
justed to a fixed setting.
0 no filter
1 GG 475 filter: to protect the patient's eye during surgery against
unnecessary (blue) radiation (retinal injury).
2 KK 40 filter: to increase color temperature
3 no filter