Technical data

20 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Xenon illumination system
1 Activating the backup lamp
The lamp module contains two xenon lamps. The second lamp is used
as a backup lamp which has to be swung into the illumination beam
path when the first lamp fails.
If the xenon lamp fails, open the lamp module as follows:
Press button (4). The lamp module is slightly ejected. Pull out the lamp
module all the way. Turn knob (1) through 180° until it snaps in. This
moves the backup lamp into the illumination beam path. Push the
lamp module all the way back into the lamp housing.
2 Yellow indicator lamp
Lights when the lamp has failed. After activation and ignition of the
backup lamp, the yellow indicator lamp goes out again.
3 Manual function
When the manual function has been activated, all electrical control
systems are inoperative. The lamp brightness is automatically ad-
justed to a fixed setting.