Technical data

32 Description
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
The S8 floor stand is optionally equipped with two different types of illumi-
nation system.
1 Halogen illumination
The halogen illumination comprises two separate systems for fiber il-
lumination. Each of the two lamp housings contains a backup lamp
which is automatically swung into the illumination beam path when the
first lamp fails.
2 Xenon illumination system
The xenon lamp generates light whose spectrum resembles that of
natural daylight. Regardless of the brightness setting, the color tem-
perature of the light always remains the same. Normal daylight film
without any additional conversion filters can therefore be used for pho-
tographic documentation. The lamp module contains two xenon
lamps. The second lamp is used as a backup lamp which has to be
manually swung into the illumination beam path when the first lamp
fails. You have to pull out the lamp module all the way before being
able to swing in the backup lamp.
Warning!
The floor stand with xenon illumination must not
be used for ophthalmic
procedures.