Technical data

36 Description
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Connection panel
1 Connector for control element
Allows the connection of a foot control panel, a hand control panel or
an operating chair with the appropriate foot switch.
2 Potential equalization bolt
3 Indicator window for rated voltage
The voltage shown here must correspond to the rated voltage avail-
able at the site of installation. You can adjust the sliding switch using
a suitable tool.
Warning!
Please observe the maximum power consumption of the two power out-
lets (4) and (5). Only connect medical devices which have been approved
by us to these outlets (4) and (5). If you use other devices, make sure that
safety is guaranteed regarding admissible ground leakage currents. The
admissible limit value of the leakage current in the stand's power cord
must not exceed 500 µA in accordance with EN60601-1/IEC 601-1. CSA
approval in compliance with UL 2601-1 only allows a maximum ground
leakage current of 300 µA.
4 Power outlet
for medical devices with a current consumption of max. 2 A.
Note:
The current of this power outlet is switched on/off using the S2 power
switch (7).
5 Power outlet
for medical devices with a current consumption of max. 5 A.
6 Power inlet
7 S2 power switch
After the stand has been switched on, the green lamp in the switch is
lit.
8 Strain relief device
The strain relief device prevents inadvertent unplugging of the fol-
lowing electrical connections:
power cable
connecting cable for foot control panel, hand control panel or op-
erating chair with an appropriate footswitch.