Technical data
38 Description
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Suspension arm
1 Locking cap of arm cover
– To open: turn 90° to the left or right.
– To close: press down and turn 90° to the left or right.
2 Adjustment screw for limiting the downward movement of the arm
Use this screw to set the minimum vertical distance (working distance)
to the surgical field. Move the surgical microscope into the working po-
sition. Turn the screw clockwise as far as it will go. Perform this setting
before
each surgical procedure.
3 Weight balancing screw
After mounting the surgical microscope including all accessories, ad-
just the balance setting of the suspension arm using this knob. The
procedure is described in the chapter “Operation“.
4 Mounting screw
for mounting the OPMI
®
coupling.
5 Locking knob
for securing the suspension arm in its horizontal position to allow
mounting of the surgical microscope. Once secured, the suspension
arm can no longer suddenly spring upward when insufficient weight is
attached.
6 Release bar
Allows non-sterile persons to release the magnetic clutches of the
ceiling mount (floor stand).
Release keys for magnetic clutches
The release keys for the magnetic clutches are located on the surgical mi-
croscope. After pressing any one of the keys, you can move the articu-
lated arm as required. When you release the key, the magnetic clutches
lock all axes simultaneously.